STATE-VARIABLE FEEDBACK CONTROL OF A MAGNETICALLY SUSPENDED CENTRIFUGAL BLOOD PUMP

Abstract

dc:description

An integral state-variable feedback controller was developed to regulate flow in a rotodynamic magnetically suspended centrifugal pump (MSCP). The controller specification was to maintain blood flow within 15% of 4 L/min against an open-loop load disturbance resulting in changes in pressure difference of ± 50 mmHg about nominal 100 mmHg occurring at 100 mmHg/min. For controller and observer design purposes, the nonlinear system was represented by a state-space model linearized about a nominal operating point with an output disturbance. A proportional integral (PI) observer was developed to estimate the unmeasurable states and disturbance. Matlab® was used for model identification and design and simulation of the controller and observer, which were then implemented using Labview®. In simulation and on the physical system, the controller maintained acceptable blood flow against an extreme unknown load disturbance occuring at a rate of approximately 200 mmHg/min.

Degree

thesis:*
Name thesis:degree_name
Master of Science
Level thesis:degree_level
masters
Discipline thesis:degree_discipline
Electrical Engineering
Grantor dc:publisher
University of Akron
Year dc:date
2007

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Selby, Normajean
Advisor dc:contributor.advisor
  • De Abreu-Garcia, Jose Alexis

Subjects

dc:subject × 9

Rights

dc:rights
Statement dc:rights
  • unrestricted
  • This thesis or dissertation is protected by copyright: all rights reserved. It may not be copied or redistributed beyond the terms of applicable copyright laws.
Language dc:language
English

Identifiers

dc:identifier.*

Chain of custody

source
Harvested from
OhioLINK
Base URL
etd.ohiolink.edu/acprod/odb_etd/ws/oai/oai
Last updated
2026-11-27
Source record
OAI-PMH GetRecord
citation

Selby, Normajean. STATE-VARIABLE FEEDBACK CONTROL OF A MAGNETICALLY SUSPENDED CENTRIFUGAL BLOOD PUMP. masters thesis, University of Akron, 2007. http://rave.ohiolink.edu/etdc/view?acc_num=akron1185567515