A Biomechanical Study of Critical Size Cranial Defect Reconstruction Techniques Using Two Bone Substitutes
Abstract
dc:descriptionWhen a skull graft is performed, a surgeon can leave the remaining defect cavity unfilled or fill it with bone cement.This research project studied cranial vault defect reconstructions immediately following surgical procedure using impact testing. This study, divided into three stages utilizing forty-two (42) fresh human cadaver specimens, addressed the clinical needs of a surgeon by evaluating three reconstruction techniques. Stage 1 compared the biomechanical characteristics of a partial thickness bone graft (n = 4) versus full thickness bone (n = 4). Stage 2 evaluated the biomechanical characteristics of a partial thickness graft filled with bone cement (n = 9) to those of an unfilled partial thickness bone graft (n = 9). Stage 3 compared the biomechanical characteristics of two bone substitutes, Stryker® HydroSet® Bone Substitute (n=6) and Norian® SRS® Fast Set Putty® (n = 6).Peak acceleration and energy absorption from initiation of impact to peak acceleration were the parameters used for analysis. In Stage 1, the parameter of peak acceleration showed a significant difference between full bone and the unfilled bone graft, full bone being favorable (p = 0.0311). The parameter of energy absorption did not show a significant difference (p = 0.1105) between the two techniques. However, the mean value of energy absorption of full bone (8.52 ± 3.88 Nm) was about two times that of partial bone (4.36 ± 2.40 Nm).In Stage 2, the parameter of peak acceleration showed a significant difference between an unfilled bone graft and a bone graft filled with Norian bone cement, the Norian sample being favorable (p = 0.0023). The parameter of energy absorption showed a significant difference in Norian's favor (p = 0.0109) between the two techniques.In Stage 3, the parameter of peak acceleration did not show a significant difference between a bone graft filled with Stryker HydroSet Bone Substitute and a bone graft filled with Norian SRS Fast Set Putty (p = 0.2343). Based on deviation from the means, Norian's performance was more predictable than Stryker's cement. The parameter of energy absorption did not show a significant difference (p = 0.5155) between the two techniques.
Degree
thesis:*- Name thesis:degree_name
- Master of Science in Engineering
- Level thesis:degree_level
- masters
- Discipline thesis:degree_discipline
- Biomedical Engineering
- Grantor dc:publisher
- University of Akron
- Year dc:date
- 2008
Author and committee
dc:creator, dc:contributor.*- Author dc:creator
-
- Porzel, Alec P.
- Advisor dc:contributor.advisor
-
- Njus, Glen
Subjects
dc:subject × 8Rights
dc:rights- Statement dc:rights
-
- unrestricted
- This thesis or dissertation is protected by copyright: all rights reserved. It may not be copied or redistributed beyond the terms of applicable copyright laws.
- Language dc:language
- English
Identifiers
dc:identifier.*- Repository record dc:identifier
- http://rave.ohiolink.edu/etdc/view?acc_num=akron1208547531