Determining Polymer Blend Surface Concentration Using Surface Layer Matrix-Assisted Laser Desorption Ionization Time of Flight Mass Spectrometry (SL-MALDI-TOF MS)
Abstract
dc:descriptionABSTRACTSurface layer matrix-assisted laser desorption ionization time-of-flight mass spectrometry (SL-MALDI-TOF MS) is a new technique to determine the surface composition of a blend film.1 The motivation of developing SL-MALDI-TOF MS is the need to measure the composition of polymer blend surfaces without having to label one component and being able to distinguish components having the same repeat, but different architectures. Polystyrene is readily ionized by MALDI making it a good candidate for this study. In this project, quantitative analyses were performed as a function of molecular weight using polystyrene standards. Such studies can be extended to similar analyses on other polymers. The first objective of this MS project was to establish the upper molecular weight (M) limit for the determination of surface composition of a polymer blend using SL-MALDI-TOF MS techniques and the variation in yield from the surface with M. We found that molecular weights from 2 to 21k could be studied. Longer polymer chains were harder to ionize, and were detected with lower efficiency. Also, the spectra of the polymer films were less resolved at higher molecular weights due to use of linear mode time-of-flight (TOF) mass analyzer and higher electronic noise. The second objective was to demonstrate the applicability of SL-MALDI-TOF MS for the measurement of surface composition of binary blends by determining the dependence of the degree of surface enrichment in isotopic blends of hydrogenous linear polystyrene (hPS) and deuterated linear polystyrene (dPS) on molecular weight. The apparent mole fraction is most accurate for blend sample with mole fraction composition near 0.5 hPS. The inaccuracy in determining the mole fraction of the surface of the film when the blend composition is far from 0.5 can be explained by the electronic noise in the spectra. The electronic noise had a very pronounced effect on the peak shape of the component with the smaller mole fraction which made the calculation of peak area inaccurate. Furthermore, SL-MALDI-TOF MS becomes less accurate as molecular weight increases. This additional loss of accuracy can be attributed to the increased difficulty of ionization and loss of resolution, affecting the ease with which peaks are distinguishable from one another.
Degree
thesis:*- Name thesis:degree_name
- Master of Science
- Level thesis:degree_level
- masters
- Discipline thesis:degree_discipline
- Polymer Science
- Grantor dc:publisher
- University of Akron
- Year dc:date
- 2014
Author and committee
dc:creator, dc:contributor.*- Author dc:creator
-
- Yao, Mengmeng
- Advisor dc:contributor.advisor
-
- Foster, Mark
Subjects
dc:subject × 4Rights
dc:rights- Statement dc:rights
-
- unrestricted
- This thesis or dissertation is protected by copyright: all rights reserved. It may not be copied or redistributed beyond the terms of applicable copyright laws.
- Language dc:language
- English
Identifiers
dc:identifier.*- Repository record dc:identifier
- http://rave.ohiolink.edu/etdc/view?acc_num=akron1407941345