Case Western Reserve University School of Graduate Studies
The design, development and implementation of electrodes used for functional electrical stimulation
Abstract
dc:descriptionElectrodes were developed and implemented in functional stimulation systems. A double helix electrode/lead, designed to be implanted without a surgical incision, was developed. The major improvement of this design over previous intramuscular electrodes was the very low axial stiffness of the electrode/lead (39 g/100% strain) allowing for stress relief during muscle contraction and increasing electrode life. Four hundred and forty-six of these electrodes were implanted in 15 paraplegic and hemiplegic subjects. Seventy percent of these electrodes continued to produce strong, stable muscle contractions after 28 months. Major causes of failure were the inability to properly place the electrode during surgery (8%), unwanted changes in muscle response (10%), electrode/lead breakage (10%), and intolerable pain (2%). A subcutaneous electrode, able to pass relatively large amounts of current (up to 100 mA @ 250 μsec pulsewidth) and designed to be implanted just below the skin with a small (1.2 cm) incision, was also developed. Three electrodes were evaluated in the lower extremity and trunk of a human subject. The muscle recruitment properties of this design were similar to recruitment properties of surface electrodes with subcutaneous electrodes using slightly less charge per stimulating pulse. This design is especially suited for use where a 'large' stimulating electric field is needed to recruit major muscle groups. 'Cylinder' and 'spiral' nerve cuff electrodes were analyzed to determine safe cuff size to avoid nerve compression. A design strategy was developed to avoid nerve damage at the edge of a nerve cuff by constructing cuffs with longitudinal 'ribs'. A 'staged' implant protocol was developed for implanting electrodes multi-channel implantable stimulator. This implant technique allows for electrode evaluation in vivo before integration with a complete stimulator system. This reduces the chance of total system failure or infection by replacing non-functioning electrodes before stimulator implantation. Characteristics which must be considered for implantable electrode design are discussed. Various strategies used to achieve desired characteristics are presented. The process used in the development of electrodes during this work was characterized and documented. A flow chart was drawn which defines the various steps required to bring an idea from initial concept to an approved design. The development process required a tedious testing and evaluation program. The complexity of development required an interdisciplinary approach to understand the nature of the problems faced and evaluate results with a...
Degree
thesis:*- Name thesis:degree_name
- Doctor of Philosophy
- Level thesis:degree_level
- doctoral
- Discipline thesis:degree_discipline
- Mechanical Engineering
- Grantor dc:publisher
- Case Western Reserve University School of Graduate Studies
- Year dc:date
- 1992
Author and committee
dc:creator, dc:contributor.*- Author dc:creator
-
- Scheiner, Avram
- Contributors dc:contributor
-
- Kicher, Thomas
Subjects
dc:subject × 1Rights
dc:rights- Statement dc:rights
-
- unrestricted
- This thesis or dissertation is protected by copyright: all rights reserved. It may not be copied or redistributed beyond the terms of applicable copyright laws.
- Language dc:language
- English
Identifiers
dc:identifier.*- Repository record dc:identifier
- http://rave.ohiolink.edu/etdc/view?acc_num=case1056388692