Quantitative and continuous measurement of cerebral blood flow by a thermal method

Abstract

dc:description

A thermal perfusion system is developed to measure cerebral blood flow continuously and quantitatively, with thermistors placed on brain surface to minimize tissue damage. Heat transfer models for tissue and self-heated thermistor are used to compare sensitivity, resolution and dynamic response under different operating modes and boundary conditions. Simulation results indicate that large perfusion sensitivity requires large thermistor while fast dynamic response calls for small thermistor. Adiabatic condition at the tissue-environment interface can maximize perfusion sensitivity and minimize thermal interference. Theoretical framework for optimal probe and system design based on measurement requirements has been established through the steady-state relations between thermal output and perfusion. The system, with lock-in amplifiers and a thermistor for reference temperature, can detect temperature signal of 0.001°C caused by perfusion changes while accommodating baseline temperature variations on the order of 0.1°C. The perfusion sensitivity and dynamic response are tested in vitro and in vivo. Noise analysis indicates that thermal fluctuations in tissue and environment are much larger than electrical noise and are the major error source. Perfusion changes associated with arterial blood pressure changes caused by bolus norepinephrine injections are used to validate the system output during continuous heating. In vivo evaluation demonstrates that the dynamic response is adequate to follow transient perfusion changes before autoregulation and perfusion cycling around 0.1 Hz. A partial analytical solution of the heat transfer model is developed, resulting consistent relations between transient and continuous heating modes. Model parameters including perfusion are estimated by nonlinear least-squares fitting from the temperature response during transient heating. A calibration coefficient is obtained from the estimated parameters without varying perfusion at different levels. Continuous perfusion changes can be quantified from the thermal output during continuous heating using the calibration coefficient. Estimated perfusion resolution of the system is about 1 mL/100g-min. Practical procedures to estimate the frequency response of thermal system have been developed for the first time. Animal experiments show that perfusion estimates from the thermal and the quantitative autoradiographic techniques are correlated. CO<sub>2</sub> reactivity test demonstrates the sensitivity of the quantitative procedure of the thermal method to perfusion changes

Degree

thesis:*
Name thesis:degree_name
Doctor of Philosophy
Level thesis:degree_level
doctoral
Discipline thesis:degree_discipline
Biomedical Engineering
Grantor dc:publisher
Case Western Reserve University School of Graduate Studies
Year dc:date
1993

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Wei, Datong
Contributors dc:contributor
  • Saidel, Gerald

Subjects

dc:subject × 3

Rights

dc:rights
Statement dc:rights
  • unrestricted
  • This thesis or dissertation is protected by copyright: all rights reserved. It may not be copied or redistributed beyond the terms of applicable copyright laws.
Language dc:language
English

Identifiers

dc:identifier.*

Chain of custody

source
Harvested from
OhioLINK
Base URL
etd.ohiolink.edu/acprod/odb_etd/ws/oai/oai
Last updated
2026-11-26
Source record
OAI-PMH GetRecord
citation

Wei, Datong. Quantitative and continuous measurement of cerebral blood flow by a thermal method. doctoral thesis, Case Western Reserve University School of Graduate Studies, 1993. http://rave.ohiolink.edu/etdc/view?acc_num=case1060354210