Abstract
dc:descriptionThis thesis details the research, design, manufacturing, and testing of an ab/adduction power unit to be used on a lower extremity hybrid neuroprosthesis (HNP) for individuals with paralysis due to spinal cord injuries. The device is intended to actuate the hip in ab/adduction for better stability and a more natural gait. The power unit is comprised of a brushless motor, two stages of compound spur gears, followed by a single planetary stage, as well as torque and position sensors. This unit builds on previous designs but differs significantly in its torque and speed requirements. Peak torque is required for up to 40% of the gait cycle, requiring more gearing (146:1) than previous power units. After bench testing, the unit was found to have a breakaway torque of 2.65-3.17 N - m, a peak torque of 65 N - m at 6.45-7.17A, and maximum speed of 137 ◦/s. It weighs 2.56 kg and has an isometric efficiency of 87% - 97.1%.
Degree
thesis:*- Name thesis:degree_name
- Master of Engineering
- Level thesis:degree_level
- masters
- Discipline thesis:degree_discipline
- EMC - Mechanical Engineering
- Grantor dc:publisher
- Case Western Reserve University School of Graduate Studies
- Year dc:date
- 2022
Author and committee
dc:creator, dc:contributor.*- Author dc:creator
-
- Jelley, Samuel Flaherty
- Chair dc:contributor.committeechair
-
- Quinn, Roger
Subjects
dc:subject × 7Rights
dc:rights- Statement dc:rights
-
- unrestricted
- This thesis or dissertation is protected by copyright: all rights reserved. It may not be copied or redistributed beyond the terms of applicable copyright laws.
- Language dc:language
- English
Identifiers
dc:identifier.*- Repository record dc:identifier
- http://rave.ohiolink.edu/etdc/view?acc_num=case1643231371698462