3D Printed Polycaprolactone Stent with Polylactic-co-glycolic Acid Nanofibrous Coating for the Treatment of Tracheal Stenosis

Abstract

dc:description

This study focused on designing and optimizing a biodegradable Polycaprolactone (PCL) stent with a hexagonal lattice structure for managing tracheal stenosis. The stent was 3D-printed and mechanically tested for stiffness, migration resistance, and long-term functionality. Results showed that the stent could effectively support the trachea and had high contractility that makes it suitable for deployment through an endotracheal tube. The stent's radial stiffness was 80% of the normal trachea’s stiffness. The anchorage force was proportional to the circumferential stiffness. The force required for initial migration was 9.3±0.03 N which was comparable to a commercial metal trachea stent (AERO® Tracheobronchial Stent) with 9±0.01 N anchorage force tested the same way. The stent was also loaded with dexamethasone and EGF for the purpose of mitigating inflammation and promoting re-epithelialization, respectively. This thesis established the stent fabrication, validated mechanical feasibility and set the groundwork for biologic delivery in preparation of future animal model.

Degree

thesis:*
Name thesis:degree_name
Master of Sciences
Level thesis:degree_level
masters
Discipline thesis:degree_discipline
EMC - Mechanical Engineering
Grantor dc:publisher
Case Western Reserve University School of Graduate Studies
Year dc:date
2024

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • FAN, DI
Advisor dc:contributor.advisor
  • Akkus, Ozan

Subjects

dc:subject × 6

Rights

dc:rights
Statement dc:rights
  • unrestricted
  • This thesis or dissertation is protected by copyright: all rights reserved. It may not be copied or redistributed beyond the terms of applicable copyright laws.
Language dc:language
English

Identifiers

dc:identifier.*

Chain of custody

source
Harvested from
OhioLINK
Base URL
etd.ohiolink.edu/acprod/odb_etd/ws/oai/oai
Last updated
2026-11-27
Source record
OAI-PMH GetRecord
citation

FAN, DI. 3D Printed Polycaprolactone Stent with Polylactic-co-glycolic Acid Nanofibrous Coating for the Treatment of Tracheal Stenosis. masters thesis, Case Western Reserve University School of Graduate Studies, 2024. http://rave.ohiolink.edu/etdc/view?acc_num=case1721304008804236