Design and Testing of a Switch Mode, Precisely Adjustable Current Driver For Use in an MRI-Compatible Robotic Catheter System

Abstract

dc:description

This thesis presents the design, implementation, and MRI performance evaluation of a novel, precisely adjustable current driver for an MRI guided robotic catheter system intended for cardiac ablation and similar procedures requiring a directed catheter in an MRI. The proposed system uses a modular, PWM-based architecture, with each channel capable of sourcing ±2 A with fine resolution, and high precision integrated current sensing for feedback. Key design considerations include RF hardening, EMI mitigation, and MRI safety through non-magnetic components, shielding, the use common mode cable traps, and low-latency controls. Extensive testing was conducted to characterize current accuracy and linearity, latency, and RF compatibility within a clinical 3 tesla MRI environment. Results demonstrate that the system achieves high linearity, low latency, and minimal interference with MRI imaging when powered from a clean DC source, validating its suitability for MRI-guided catheter steering. This work builds upon prior systems for robotically steered MRI guided catheters and introduces improvements in size, efficiency, MRI compatibility and control architecture, demonstrating the potential for a switch mode, digitally controlled current device to function in a highly RF sensitive environment and laying the foundation for future integration into interventional MRI procedures.

Degree

thesis:*
Name thesis:degree_name
Master of Engineering
Level thesis:degree_level
masters
Discipline thesis:degree_discipline
EECS - Electrical Engineering
Grantor dc:publisher
Case Western Reserve University School of Graduate Studies
Year dc:date
2026

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Cowles, Tobias
Advisor dc:contributor.advisor
  • Çavuşoğlu, Cenk

Subjects

dc:subject × 2

Rights

dc:rights
Statement dc:rights
  • restricted--full text unavailable until 2028-05-15
  • This thesis or dissertation is protected by copyright: all rights reserved. It may not be copied or redistributed beyond the terms of applicable copyright laws.
Language dc:language
English

Identifiers

dc:identifier.*

Chain of custody

source
Harvested from
OhioLINK
Base URL
etd.ohiolink.edu/acprod/odb_etd/ws/oai/oai
Last updated
2026-11-27
Source record
OAI-PMH GetRecord
citation

Cowles, Tobias. Design and Testing of a Switch Mode, Precisely Adjustable Current Driver For Use in an MRI-Compatible Robotic Catheter System. masters thesis, Case Western Reserve University School of Graduate Studies, 2026. http://rave.ohiolink.edu/etdc/view?acc_num=case1778112530641885