A CFD Model of Mixing in a Microfluidic Device for Space Medicine Technology

Abstract

dc:description

The DNA Medicine Institute (DMI) is currently developing a device to be used for blood analysis to satisfy the unique requirements of space medicine applications. A key component of that device is the micromixer, which will ensure mixing and dilution of reagents utilized for detection assays. As part of the device design process, the micromixer was modeled, and the mixing characteristics were analyzed and compared to experimental data. The experimental data was based on a top-view of the system and, lacking data throughout the fluid domain, could not provide the insight into the mixing process that modeling could readily provide. COMSOL, a Finite Element Method (FEM) package, was used to model the mixer. The mixer design is essentially a spiral channel and relies on centrifugal effects, or Dean flow forces that arise from flows in curved channels, to enhance mixing. A computational model of DMI’s spiral mixer was analyzed and compared to experimental data for flow ranging in Reynolds number between 8 and 90. The Dean number range was between 0 and 25. The fluids modeled were miscible and Newtonian. It was observed that at Reynolds number less than 12 (De < 2), the mixing occurred primarily by diffusion and at Reynolds numbers above 30 (De > 11), convective forces dominated. In an intermediate range, Reynolds numbers between 12 and 30 (De 2 – 11), mixing appeared to be enhanced as both diffusion and convection aided the mixing. Due to the rotational nature of the flow, this was not readily apparent from the experimental data. The model is a good tool to optimize design choices since the numerical data can be used to quantify mixing characteristics throughout the entire mixer volume, thereby providing a better insight into mixing performance.

Degree

thesis:*
Name thesis:degree_name
Master of Science in Mechanical Engineering
Level thesis:degree_level
masters
Discipline thesis:degree_discipline
Fenn College of Engineering
Grantor dc:publisher
Cleveland State University
Year dc:date
2011

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • McKay, Terri L.
Contributors dc:contributor
  • Ibrahim, Mounir
  • Fodor, Petru
  • Kaufman, Miron
  • Nelson, Emily

Subjects

dc:subject × 7

Rights

dc:rights
Statement dc:rights
  • unrestricted
  • This thesis or dissertation is protected by copyright: all rights reserved. It may not be copied or redistributed beyond the terms of applicable copyright laws.
Language dc:language
English

Identifiers

dc:identifier.*

Chain of custody

source
Harvested from
OhioLINK
Base URL
etd.ohiolink.edu/acprod/odb_etd/ws/oai/oai
Last updated
2026-11-26
Source record
OAI-PMH GetRecord
citation

McKay, Terri L.. A CFD Model of Mixing in a Microfluidic Device for Space Medicine Technology. masters thesis, Cleveland State University, 2011. http://rave.ohiolink.edu/etdc/view?acc_num=csu1305563573