Back to search

University of Toledo Health Science Campus

Dosimetric Verification of the ADAC Pinnacle3 Pencil Beam Algorithm For Clinical Electrons In Presence of Cerrobend Blocking

Abstract

dc:description

The purpose of this study was to commission planning electrons in the ADAC Pinnacle3treatment planning system and to verify its accuracy with clinical Cerrobend blockedfields. Pinnacle utilizes the Hogstrom electron pencil beam algorithm to perform itselectron dose calculation. Complete set of measurement beam data in water phantom wasfirst acquired and verified for consistency. The data was then entered into the planningsystem to develop a measurement and parameter-based model per beam. After thesystem was commissioned for clinical use, additional verification measurements wereperformed with unique field blocking. Comparisons between measured and computeddose from Pinnacle demonstrated the calculation was accurate and within clinicaltolerance for all field blocking at standard treatment distance. For extended distance,relative depth dose prediction was poor only in the case with very high percentage fieldblocking. Relative cross-beam dose results were also poorer but were still acceptable forclinical use.

Degree

thesis:*
Name thesis:degree_name
Master of Science in Biomedical Sciences (MSBS)
Level thesis:degree_level
masters
Discipline thesis:degree_discipline
College of Graduate Studies
Grantor dc:publisher
University of Toledo Health Science Campus
Year dc:date
2007

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Chan, Philip
Chair dc:contributor.committeechair
  • Parsai, E. Ishmael

Subjects

dc:subject × 5

Rights

dc:rights
Statement dc:rights
  • unrestricted
  • This thesis or dissertation is protected by copyright: all rights reserved. It may not be copied or redistributed beyond the terms of applicable copyright laws.
Language dc:language
English

Identifiers

dc:identifier.*

Chain of custody

source
Harvested from
OhioLINK
Base URL
etd.ohiolink.edu/acprod/odb_etd/ws/oai/oai
Last updated
2026-11-27
Source record
OAI-PMH GetRecord
citation

Chan, Philip. Dosimetric Verification of the ADAC Pinnacle3 Pencil Beam Algorithm For Clinical Electrons In Presence of Cerrobend Blocking. masters thesis, University of Toledo Health Science Campus, 2007. http://rave.ohiolink.edu/etdc/view?acc_num=mco1211993729