University of Toledo Health Science Campus
Dosimetric Verification of the ADAC Pinnacle3 Pencil Beam Algorithm For Clinical Electrons In Presence of Cerrobend Blocking
Abstract
dc:descriptionThe purpose of this study was to commission planning electrons in the ADAC Pinnacle3treatment planning system and to verify its accuracy with clinical Cerrobend blockedfields. Pinnacle utilizes the Hogstrom electron pencil beam algorithm to perform itselectron dose calculation. Complete set of measurement beam data in water phantom wasfirst acquired and verified for consistency. The data was then entered into the planningsystem to develop a measurement and parameter-based model per beam. After thesystem was commissioned for clinical use, additional verification measurements wereperformed with unique field blocking. Comparisons between measured and computeddose from Pinnacle demonstrated the calculation was accurate and within clinicaltolerance for all field blocking at standard treatment distance. For extended distance,relative depth dose prediction was poor only in the case with very high percentage fieldblocking. Relative cross-beam dose results were also poorer but were still acceptable forclinical use.
Degree
thesis:*- Name thesis:degree_name
- Master of Science in Biomedical Sciences (MSBS)
- Level thesis:degree_level
- masters
- Discipline thesis:degree_discipline
- College of Graduate Studies
- Grantor dc:publisher
- University of Toledo Health Science Campus
- Year dc:date
- 2007
Author and committee
dc:creator, dc:contributor.*- Author dc:creator
-
- Chan, Philip
- Chair dc:contributor.committeechair
-
- Parsai, E. Ishmael
Subjects
dc:subject × 5Rights
dc:rights- Statement dc:rights
-
- unrestricted
- This thesis or dissertation is protected by copyright: all rights reserved. It may not be copied or redistributed beyond the terms of applicable copyright laws.
- Language dc:language
- English
Identifiers
dc:identifier.*- Repository record dc:identifier
- http://rave.ohiolink.edu/etdc/view?acc_num=mco1211993729