Dosimetric Evaluation of Three Partial Breast Irradiation Devices and the Dosimetric Effect of Tissue Thickness Surrounding a Multi-Lumen Partial Breast Applicator

Abstract

dc:description

<p>Many High Dose Rate treatment planning systems that are in use fail to correct for heterogeneities. If the treatment planning system does not correct for heterogeneities, it would assume that the patient is receiving full scatter when in reality, the patient will possibly be underdosed. A 1cm diameter planning target volume for a lumpectomy cavity could extend beyond the skin or chest wall for the patient and could be a great problem when it comes to treatment with the MammoSite® single lumen breast applicator. A previous Monte Carlo study tested 3 MammoSite® balloon sizes at various depths beyond the planning target volume to see how much tissue would be needed to achieve full scatter. The results showed that on average, if there was no tissue beyond the prescription line of 1cm there would be a 10% dose reduction for the breast – skin interface.</p><p>The purpose of this study is to use the Strut Adjusted Volume Implant (SAVI) multi-lumen breast applicator to re-create the measurements done with the MammoSite® balloon and expand these measurements to include tissue thicknesses less than the PTV. Previous simulations with the MammoSite® were done using Monte Carlo, with tissue thicknesses beyond the planning target volume of 0 – 10cm. This study will re-create these measurements using Metal Oxide Semiconductor Field Effect Transistors (MOSFETs) and also take measurements below the prescription line of 1cm due to the ability of the SAVI applicator to adjust dose to the skin.</p>

Degree

thesis:*
Name thesis:degree_name
Master of Science in Biomedical Sciences (MSBS)
Level thesis:degree_level
masters
Discipline thesis:degree_discipline
College of Medicine
Grantor dc:publisher
University of Toledo Health Science Campus
Year dc:date
2010

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Detwiler, Jordyn A.
Contributors dc:contributor
  • David, Pearson

Subjects

dc:subject × 3

Rights

dc:rights
Statement dc:rights
  • unrestricted
  • This thesis or dissertation is protected by copyright: all rights reserved. It may not be copied or redistributed beyond the terms of applicable copyright laws.
Language dc:language
English

Identifiers

dc:identifier.*

Chain of custody

source
Harvested from
OhioLINK
Base URL
etd.ohiolink.edu/acprod/odb_etd/ws/oai/oai
Last updated
2026-11-26
Source record
OAI-PMH GetRecord
citation

Detwiler, Jordyn A.. Dosimetric Evaluation of Three Partial Breast Irradiation Devices and the Dosimetric Effect of Tissue Thickness Surrounding a Multi-Lumen Partial Breast Applicator. masters thesis, University of Toledo Health Science Campus, 2010. http://rave.ohiolink.edu/etdc/view?acc_num=mco1286688741