<p> The role of the dose-volume histogram (DVH) is rapidly expanding in radiation oncology treatment planning. DVHs are already relied upon to differentiate between two similar plans and evaluate organ-at-risk dosage. Their role will become even more important as progress continues towards implementing biologically based treatment planning systems. Therefore it is imperative that the accuracy of DVHs is evaluated and reappraised after any major software or hardware upgrades, affecting a treatment planning system (TPS). The purpose of this work is to create and implement a comprehensive quality assurance procedure evaluating dose volume histograms to insure their accuracy while satisfying American College of Radiology guidelines.<p><p> Virtual phantoms of known volumes were created in Pinnacle TPS and exposed to different beam arrangements. Variables including grid size and slice thickness were varied and their effects were analyzed. The resulting DVHs were evaluated by comparison to the commissioned percent depth dose values using a custom Excel spreadsheet. After determining the uncertainty of the DVH based on these variables, multiple second check calculations were performed using MIM Maestro and Matlab software packages.<p><p> The uncertainties of the DVHs were shown to be less than &#xb1; 3%. The average uncertainty was shown to be less than &#xb1; 1%. The second check procedures resulted in mean percent differences less than 1% which confirms the accuracy of DVH calculation in Pinnacle and the effectiveness of the quality assurance template.<p><p> The importance of knowing the limits of accuracy of the DVHs, which are routinely used to assess the quality of clinical treatment plans, cannot be overestimated. The developed comprehensive QA procedure evaluating the accuracy of the DVH statistical analysis will become a part of our clinical arsenal for periodic tests of the treatment planning system. It will also be performed at the time of commissioning and after any major software/hardware upgrades.<p>

Degree

thesis:*
Name thesis:degree_name
Master of Science in Biomedical Sciences (MSBS)
Level thesis:degree_level
masters
Discipline thesis:degree_discipline
College of Medicine
Grantor dc:publisher
University of Toledo Health Science Campus
Year dc:date
2013

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Davenport, David Alan
Chair dc:contributor.committeechair
  • Shvydka, Diana

Subjects

dc:subject × 7

Rights

dc:rights
Statement dc:rights
  • unrestricted
  • This thesis or dissertation is protected by copyright: all rights reserved. It may not be copied or redistributed beyond the terms of applicable copyright laws.
Language dc:language
English

Identifiers

dc:identifier.*

Chain of custody

source
Harvested from
OhioLINK
Base URL
etd.ohiolink.edu/acprod/odb_etd/ws/oai/oai
Last updated
2026-11-27
Source record
OAI-PMH GetRecord
citation

Davenport, David Alan. Development of a Quality Assurance Procedure for Dose Volume Histogram Analysis. masters thesis, University of Toledo Health Science Campus, 2013. http://rave.ohiolink.edu/etdc/view?acc_num=mco1372672842