This dissertation reports the development of novel implantable oxygen sensors made by the encapsulation of particulate electron paramagnetic resonance (EPR) oximetry probes in a biocompatible polymer matrix. Implants were fabricated by cast molding and polymerization, and exhibited a stable and linear oxygen calibration. In vitro testing of these polymeric biomaterials established their robustness and biostability. In addition, when cultured with cells (in the medium), the implants exhibited excellent in vitro biocompatibility and biodurability. In vivo testing of the long-term oxygen sensing performance showed stable, real-time response for up to 70 days. Functional validation using clinically relevant animal models of disease revealed that the implants were capable of accurately reporting the oxygenation of the tissue under study. The development of implantable oxygen sensors, presented in this dissertation, will enable the widespread application of EPR for in vivo oximetry and the successful clinical transformation of this promising research technique.

Degree

thesis:*
Name thesis:degree_name
Doctor of Philosophy
Level thesis:degree_level
doctoral
Discipline thesis:degree_discipline
Biomedical Engineering
Grantor dc:publisher
The Ohio State University
Year dc:date
2008

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Meenakshisundaram, Guruguhan
Advisor dc:contributor.advisor
  • Lee, Stephen

Subjects

dc:subject × 9

Rights

dc:rights
Statement dc:rights
  • unrestricted
  • This thesis or dissertation is protected by copyright: all rights reserved. It may not be copied or redistributed beyond the terms of applicable copyright laws.
Language dc:language
English

Identifiers

dc:identifier.*

Chain of custody

source
Harvested from
OhioLINK
Base URL
etd.ohiolink.edu/acprod/odb_etd/ws/oai/oai
Last updated
2026-11-27
Source record
OAI-PMH GetRecord
citation

Meenakshisundaram, Guruguhan. Development of novel implantable sensors for biomedical oximetry. doctoral thesis, The Ohio State University, 2008. http://rave.ohiolink.edu/etdc/view?acc_num=osu1217427728