Back to search

The Ohio State University

Evaluation of efficacy and biocompatibility of indirect intraocular pressure monitoring using a telemetric scleral sensor

Abstract

dc:description

Purpose. To evaluate efficacy, biocompatibility and stability of a surgically implanted telemetric scleral sensor (TSS). To obtain an intraocular pressure (IOP) measurement without the need for topical anesthesia, patient compliance or corneal contact as is necessary with conventional applanation or rebound tonometry. Methods. A TSS was surgically implanted subconjunctivally and sutured to the sclera of the right eye of 10 juvenile female NZW rabbits. IOP was monitored using the TSS and a Tono-Vet twice daily for two weeks, then twice weekly until sacrifice. Ophthalmic examination was performed weekly and the eyes were graded according to the modified Hackett-MacDonald scheme. Two rabbits were sacrificed at 2, 4, 8, 12 and 16 weeks. Manometric IOP values were also recorded in 5mmHg intervals up to 70mmHg and compared to TSS. Eyes were submitted for histopathology. Results. Fibrosis and inflammation between the sclera and implant were graded on a scale of 0-4 for extent and severity, respectively. The grade of fibrosis and inflammatory response ranged from 1-4 (mean 2.5) and 0-2.3 (mean 1.2), respectively and was not associated with time. The predominant cell types recorded were the granulocyte and lymphocyte. Scleral thinning was observed at the site of TSS contact. Initial readings showed good correlation between the TSS and Tono-Vet, but premature dislocation of the TSS in 6 rabbits limited collection of manometric data. Conclusions. Although initial readings showed good correlation between the TSS and Tono-Vet, the overall size and design of the TSS contributed to fibrosis and premature dislocation of the sensor.

Degree

thesis:*
Name thesis:degree_name
Master of Science
Level thesis:degree_level
masters
Discipline thesis:degree_discipline
Veterinary Clinical Sciences
Grantor dc:publisher
The Ohio State University
Year dc:date
2010

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Robinson, Christa Lee
Advisor dc:contributor.advisor
  • Wilkie, David

Subjects

dc:subject × 7

Rights

dc:rights
Statement dc:rights
  • unrestricted
  • This thesis or dissertation is protected by copyright: all rights reserved. It may not be copied or redistributed beyond the terms of applicable copyright laws.
Language dc:language
English

Identifiers

dc:identifier.*

Chain of custody

source
Harvested from
OhioLINK
Base URL
etd.ohiolink.edu/acprod/odb_etd/ws/oai/oai
Last updated
2026-11-27
Source record
OAI-PMH GetRecord
citation

Robinson, Christa Lee. Evaluation of efficacy and biocompatibility of indirect intraocular pressure monitoring using a telemetric scleral sensor. masters thesis, The Ohio State University, 2010. http://rave.ohiolink.edu/etdc/view?acc_num=osu1274142444