A comparison of three methods of measuring central corneal thickness in normal and thinned corneas
Abstract
dc:description<p>The thickness of the cornea is an important parameter in Goldmann applanation tonometry, corneal disease, contact lens wear, and refractive surgery. Pachymetry is the measurement of corneal thickness, and currently ultrasound pachymetry is the gold standard. Several non-contact instruments, the Orbscan II (Orbtek, Inc.) and the Visante anterior segment-optical coherence tomographer (Carl Zeiss Meditec) have been introduced that provide more information about the cornea including corneal thickness topography maps. </p><p>The goal of the present study was to determine repeatability of the instruments, to assess agreement between the instruments, and to assess the effect of corneal thickness on repeatability and agreement. Because each instrument is based on different principles, measurements may be biased when normal versus abnormally thin corneas are measured.</p><p>Forty-five subjects (15 normal, 15 post-LASIK, and 15 keratoconus) were examined on two occasions, separated by one-to-seven days. Three measurements (later averaged) were made on the right eye of each subject using the three techniques. The subjects were analyzed as three groups: normal corneas, post-LASIK and keratoconus. Mixed model analysis of variance was used to compare between instruments, visit, and within groups. For each technique and cornea type, the difference between instruments and visits was determined for each subject and analyzed using Bland-Altman analysis to produce the mean, standard deviation, and 95% limits of agreement (LoA). A paired two-sample t test was also conducted for validity and repeatability. </p><p>For normal corneas, mean central corneal thickness was as follows: 550 ± 34 µm for ultrasound, 541 ± 44 µm for Orbscan II, and 532 ± 38 µm for Visante. For post-LASIK corneas, mean central corneal thickness was as follows: 502 ± 40 µm for ultrasound, 495 ± 52 µm for Orbscan II, and 485 ± 42 µm for Visante. For keratoconic corneas mean central corneal thickness was as follows: 529 ± 46 µm for ultrasound, 498 ± 44 µm for Visante, and 490 ± 53 µm for Orbscan II. Mixed model analysis showed a significant interaction between instrument and cornea types (p=0.012). That is, the instruments not only measured significantly different from each other (p < 0.001) but also significantly different when measuring the various cornea types. Repeatability was poorer for all techniques in the keratoconic corneas and for ultrasound. For normal corneas, the Visante was the most repeatable: 95% LoA of –8 to +8 µm followed by the Orbscan: 95% LoA of –11 to +12 um, and ultrasound: 95% LoA of –15 to +19 um. For post-LASIK corneas, the Orbscan was the most repeatable: 95% LoA –11 to +11 µm followed by the Visante: 95% LoA of –10 to +13, and ultrasound: 95% LoA –24 to +16 um. For keratoconic corneas, the Visante was the most repeatable: 95% LoA –36 to +37 µm followed by the Orbscan: 95% LoA –31 to +47 µm and ultrasound: 95% LoA –40 to +53 um.</p>
Degree
thesis:*- Name thesis:degree_name
- Master of Science
- Level thesis:degree_level
- masters
- Discipline thesis:degree_discipline
- Vision Science
- Grantor dc:publisher
- The Ohio State University
- Year dc:date
- 2010
Author and committee
dc:creator, dc:contributor.*- Author dc:creator
-
- Colling, Amber J.
- Advisor dc:contributor.advisor
-
- Bullimore, Mark
Subjects
dc:subject × 5Rights
dc:rights- Statement dc:rights
-
- unrestricted
- This thesis or dissertation is protected by copyright: all rights reserved. It may not be copied or redistributed beyond the terms of applicable copyright laws.
- Language dc:language
- English
Identifiers
dc:identifier.*- Repository record dc:identifier
- http://rave.ohiolink.edu/etdc/view?acc_num=osu1275058650