Back to search

The Ohio State University

Design and Fabrication of an Electromagnetic Probe for Biomedical Applications

Abstract

dc:description

<p>This research entails the design, fabrication, and characterization of an electromagnetic (EM) probe capable of detecting variations in EM properties in conducting media by generating eddy currents in the sample. A targeted application of this method is to distinguish cancer from normal tissue, in hopes eventually to serve as a real-time, intraoperative detection tool for surgeons in the operating room (OR). The probe, similar to a transformer, consists of a primary coil and secondary (detector) coil, inductively coupled by virtue of a time varying current applied to the primary coil. This results in a periodic detector coil voltage trace comprising several peaks. When a conductive sample such as tissue is placed in close proximity to the probe, changes in magnitude and temporal occurrence of these peaks are observed. The shifts observed in the detector coil voltage peaks are affected by the path the eddy current travels in the conducting specimen and by the magnitude of the eddy current itself.</p><p>This work differs from previous work in two principal aspects. First, the prototype probe known in this work as B1 is hardwired to a printed circuit board, as opposed to being non-permanently connected on a breadboard. Experimental measurements show that differences between the permanent and non-permanent configurations are negligible. A second aspect is the design methodology used to make B1 smaller than any of its predecessors, roughly on the order of millimeters in diameter.In making a smaller probe, each coil’s self inductance is inherently smaller; however, by placing additional capacitors in parallel with each coil to increase the net capacitance, a temporal response mimicking that of its larger predecessor can be obtained provided the characteristic LC response is maintained. The experimental measurements and numerical simulations discussed in this thesis demonstrate the significance of the circuit element properties of the probe and how these can be used to model the probe and predict its basic operating characteristics.</p><p>The drive toward miniaturization stems from the need to have ever increasing spatial resolution. Cancer cells are typically on the order of 20 μm in size, and cell-level detection in tissues is the ultimate goal. Although the prototype probe developed in this work is incapable of resolving spatial scales as small as a cell, the design methodology developed here shows promise for enabling the design of probes capable of resolution of 1 mm or smaller. The improved resolution of B1 over its predecessor prototypes is demonstrated in imaging experiments on a paraffin wax phantom with embedded copper rings simulating tumors.</p>

Degree

thesis:*
Name thesis:degree_name
Master of Science
Level thesis:degree_level
masters
Discipline thesis:degree_discipline
Mechanical Engineering
Grantor dc:publisher
The Ohio State University
Year dc:date
2011

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Wilson, Michelle Lynn
Advisor dc:contributor.advisor
  • Subramaniam, Vish

Subjects

dc:subject × 3

Rights

dc:rights
Statement dc:rights
  • unrestricted
  • This thesis or dissertation is protected by copyright: all rights reserved. It may not be copied or redistributed beyond the terms of applicable copyright laws.
Language dc:language
English

Identifiers

dc:identifier.*

Chain of custody

source
Harvested from
OhioLINK
Base URL
etd.ohiolink.edu/acprod/odb_etd/ws/oai/oai
Last updated
2026-11-27
Source record
OAI-PMH GetRecord
citation

Wilson, Michelle Lynn. Design and Fabrication of an Electromagnetic Probe for Biomedical Applications. masters thesis, The Ohio State University, 2011. http://rave.ohiolink.edu/etdc/view?acc_num=osu1313509778