A Comparison of Gas Flow Resistane in Parker Flex-tip and Mallinckrodt RAE Nasal Endotracheal Tubes

Abstract

dc:description

Introduction: The newer Parker Flex-Tip endotracheal tube causes less trauma and/or bleeding than more traditional designs during naso-tracheal intubation. However, the Parker tube appears to deform to a greater extent during normal use than other endotracheal tubes. This study aimed to investigate the resistance to gas flow that both the Mallinckrodt RAE and Parker Flex-Tip demonstrate, and how these two types of tubes may differ in their ability to withstand the deformation caused by thermosoftening and clinical positioning during normal use. Methods: Mallinckrodt RAE (Mallinckrodt Medical, Dublin, Ireland) and Parker Flex-Tip (Parker Medical, Highlands Ranch, CO) cuffed nasal endotracheal tubes, sizes 4.0, 4.5, 5.0, 5.5, 6.0, 7.0, 7.5 were connected to two Certifier FA Plus high flow modules (TSI Incorporated, Shoreview, MN), one connected to the proximal and the second, via a 90° fitting, to the distal end of each tube. A precise gas flow of 6 L/min was confirmed and measurements of pressure were collected at both ends of each tube as each tube was bent to angles of 70, 60, 50, and 45 degrees. Statistical analysis was done using a repeated measures analysis of variance and making a Satterthwaite adjustment to degrees of freedom to eliminate any bias due to heteroscedasticity. The alpha level was set at 0.05. The three main effects found were the brand of the tube, the internal diameter of the tube, and the degree of bend of the tube. Results: None of the two-way interactions tested was found to be statistically significant (P=0.198) (Table 1). However, the Parker Flex-tip nasal endotracheal tube, when compared to the Mallinckrodt RAE tube, demonstrated a statistically significant higher resistance to gas flow, at 98 degrees Fahrenheit, over the entire range of tube diameters tested at both 50 degree and 60 degree bends (P=0.013) (Table 2) (Figure 4). Conclusion: As is to be expected, smaller diameter nasal endotracheal tubes are associated with an increased resistance to gas flow. Interestingly, the resistance to gas flow was shown to be greater in the Parker Flex-tip nasal endotracheal tube under the conditions of our testing. When used properly, the Parker Flex-tip nasal endotracheal tube has been shown to cause less trauma and epistaxis during routine naso-tracheal intubation. However, the increased resistance to gas flow may have a negative impact on the physiologic work of breathing in nasally intubated patients; particularly in the smaller size tubes used in the pediatric population. The physiological impact of this in terms of the increased work of breathing that may be required through a Parker tube warrants further investigation.

Degree

thesis:*
Name thesis:degree_name
Master of Science
Level thesis:degree_level
masters
Discipline thesis:degree_discipline
Dentistry
Grantor dc:publisher
The Ohio State University
Year dc:date
2013

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Perry, Joshua L.
Advisor dc:contributor.advisor
  • Prior, Simon

Subjects

dc:subject × 5

Rights

dc:rights
Statement dc:rights
  • unrestricted
  • This thesis or dissertation is protected by copyright: all rights reserved. It may not be copied or redistributed beyond the terms of applicable copyright laws.
Language dc:language
English

Identifiers

dc:identifier.*

Chain of custody

source
Harvested from
OhioLINK
Base URL
etd.ohiolink.edu/acprod/odb_etd/ws/oai/oai
Last updated
2026-11-27
Source record
OAI-PMH GetRecord
citation

Perry, Joshua L.. A Comparison of Gas Flow Resistane in Parker Flex-tip and Mallinckrodt RAE Nasal Endotracheal Tubes. masters thesis, The Ohio State University, 2013. http://rave.ohiolink.edu/etdc/view?acc_num=osu1380284347