Strain analysis and fracture strength of different abutments for cement-retained crowns on an internal hexagon implant
Abstract
dc:descriptionPurpose: Many aftermarket abutments for cement-retained crowns are available for the Tapered Screw-Vent implant (Zimmer Dental, Carlsbad, CA, USA). Of these aftermarket abutments this is wide variation, from stock to custom abutments, as well as material differences such as zirconia, titanium, or a combination of the two. This study was performed in two parts: the aim of Part I was to measure and compare different strains conferred to the bone around the implants for nine different abutments for cement-retained crowns on an implant with an internal hexagon platform. The aim of Part II was to measure and compare the force required to cause fracture of ten different abutments for cement-retained crowns on an implant with an internal hexagon platform. Material and Methods: Nine 4.1mm x 11.5mm Tapered Screw-Vent implants were placed into a 12 x 2 x 8mm resin block for strain measurements. Five abutment specimens of each of the nine different abutments (n = 44 total) were tried into one of the nine implants. Monolithic zirconia crowns were then fabricated for each of the nine different abutments and the crowns were cyclically loaded at thirty degrees two times at a frequency of 2Hz and strain was measured and recorded. Ten Tapered Screw-Vent Implants were then individually secured in a loading apparatus, and three abutment specimens of each of the ten different abutments (n = 30 total) were loaded at a 30 degree angle until fracture of the implant abutment complex was achieved. The strain to the resin block was determined using three-dimensional digital image correlation (3D DIC), an optical measurement technique. Commercial image correlation software (VIC 3D v7.0 , Correlated Solutions, Inc. Columbia, SC, USA) was used to analyze the strain around the implants. Data for maximal and minimal principle strains as well as load to fracture were compared using analysis of variance with a Tukey-Kramer post-hoc test (alpha = .05). Results: Strain measurements showed no significant difference between any of the abutments when looking at minimal (compression) principle strains (p>.05). For maximal (tensile) principle strains the Atlantis Zirconia abutment and the Zimmer Patient Specific Abutment showed the highest and second-highest strain around the implant, respectively, with the Atlantis Zirconia being significantly greater than all abutments excluding the Zimmer PSA, and the PSA being significantly greater than the AstraTech ZirDesign and Legacy Straight Contoured abutment in titanium as well as zirconia (p<.05). For load to fracture tests, the Inclusive Custom Abutment in Zirconia and AstraTech ZirDesign zirconia abutment showed the lowest and second lowest load to fracture, respectively, which were significantly lower than all other abutments (p<.05). The highest overall fracture strength were the Legacy Straight Contoured Abutment in Titanium and Legacy Straight Contoured Abutment in Zirconia, which were significantly greater than all other abutments (p<.05). The Zimmer PSA abutment was the only abutment that showed no fracture of any of the components. Overall the full contour zirconia abutments fractured at an average of 278.03N vs 589.69N load to fracture of all ten abutments and 699.32N load to fracture of titanium abutments. Conclusion: When selecting an abutment for a cement-retained crown on a Tapered Screw-Vent implant, the material selection of full contour zirconia vs titanium as well as off-market brands vs. the implant brand being used should be considered heavily. When comparing the strain to the implants, the Zimmer PSA and Atlantis Zirconia abutments conferred the most tensile strain. For load to fracture the AstraTech ZirDesign and Inclusive Custom Implant Abutment in zirconia showed significantly less fracture resistance than any of the other implants. For zimmer vs. aftermarket brands, the Zimmer PSA abutment was the only abutment that had no fracture of any of the components. The Zimmer Contour Abutment in Zirconia had a significantly greater fracture resistance than the full zirconia abutments without titanium components.
Degree
thesis:*- Name thesis:degree_name
- Master of Science
- Level thesis:degree_level
- masters
- Discipline thesis:degree_discipline
- Dentistry
- Grantor dc:publisher
- The Ohio State University
- Year dc:date
- 2014
Author and committee
dc:creator, dc:contributor.*- Author dc:creator
-
- Salaita, Louai Ghaleb
- Advisor dc:contributor.advisor
-
- Yilmaz, Burak
Subjects
dc:subject × 8Rights
dc:rights- Statement dc:rights
-
- unrestricted
- This thesis or dissertation is protected by copyright: some rights reserved. It is licensed for use under a Creative Commons license. Specific terms and permissions are available from this document's record in the OhioLINK ETD Center.
- Language dc:language
- English
Identifiers
dc:identifier.*- Repository record dc:identifier
- http://rave.ohiolink.edu/etdc/view?acc_num=osu1404820351