Comparing Cone Beam Computed Tomography with Multi-Slice Computed Tomography in Diagnosing Osseous Defects at the Mandibular Condyle

Abstract

dc:description

Objectives: Previous studies indicated that orthodontic-grade CBCTs are limited in displaying small defects at the mandibular condyles for diagnosis. This study investigated whether this limitation was inherent to CBCT by comparing CBCT with multislice CT, and whether it can be overcome by image segmentation and color mapping. Methods: Nine fresh pig heads (18 condyles, 36 medial/lateral condylar regions) were used. Small osseous defects (diameter and depth =1.5mm) were created at condylar, medial and lateral, regions demarcated by gutta percha markers. After restoring the overlying soft tissues, the pig heads underwent orthodontic-grade CBCT scans (i-CAT, 0.4mm voxel-size) and medical-grade CT scans (GE LightSpeed, 0.625mm voxel-size) scans. Subsequently, two calibrated and blinded raters diagnosed defect number in each condylar region from CBCT and CT images using Dolphin-3D software without image segmentation, then 1-week later with proprietary image segmentation and color mapping tools of Dolphin-3D. Condylar PVS impressions were collected and evaluated by the same raters to obtain physical diagnosis. Re-diagnosis was made on randomly selected subsamples to assess reliability. Using physical diagnosis as references, the accuracy of imaging diagnosis were assessed and statistically compared among varied imaging /analysis methods. Results: Image diagnoses of all imaging/analysis methods showed good or excellent intra- and inter-rater reliability except for those of segmented CBCT images, which were substantially lower. The numbers of over- and under-diagnosis per condylar region were not significantly differences among varied imaging/analysis methods (Wilcoxon tests, p>0.05) but those from CT tended to be lower than CBCT. Classification functions demonstrate substantial lower sensitivity and accuracy with CBCT than with CT. Logistic regression showed CT had a significantly higher probability (odds ratio 2.4) than CBCT in reaching correct diagnosis, while the use of image segmentation and color mapping did not improve the diagnostic accuracy from CBCT images. Conclusions: Even at a lower voxel-size than medical CT, orthodontic-grade CBCT images of mandibular condyles are inherently more difficult than CT images for diagnosing small condylar defects, and this limitation cannot be overcome by using image segmentation and color mapping.

Degree

thesis:*
Name thesis:degree_name
Master of Science
Level thesis:degree_level
masters
Discipline thesis:degree_discipline
Dentistry
Grantor dc:publisher
The Ohio State University
Year dc:date
2015

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Jones, Elizabeth Marie Matteson
Advisor dc:contributor.advisor
  • Sun, Zongyang

Subjects

dc:subject × 1

Rights

dc:rights
Statement dc:rights
  • unrestricted
  • This thesis or dissertation is protected by copyright: all rights reserved. It may not be copied or redistributed beyond the terms of applicable copyright laws.
Language dc:language
English

Identifiers

dc:identifier.*

Chain of custody

source
Harvested from
OhioLINK
Base URL
etd.ohiolink.edu/acprod/odb_etd/ws/oai/oai
Last updated
2026-11-27
Source record
OAI-PMH GetRecord
related terms
citation

Jones, Elizabeth Marie Matteson. Comparing Cone Beam Computed Tomography with Multi-Slice Computed Tomography in Diagnosing Osseous Defects at the Mandibular Condyle. masters thesis, The Ohio State University, 2015. http://rave.ohiolink.edu/etdc/view?acc_num=osu1427278935