Back to search

The Ohio State University

The Effect of Cyclic Loading on the Zirconia/Titanium Implant Abutment Interface and the Mechanism of Failure between Three Different Types of Abutments

Abstract

dc:description

A single implant restoration in the anterior portion of the mouth poses many challenges. With the increasing demand for ideal esthetics, masking the greyish discoloration from the titanium abutment is very difficult especially in patients with a high smile line or a thin gingival biotype. The one-piece zirconia abutment introduced, however was shown to have many clinical complications such as fracture of the zirconia connection and wear of the titanium from the implant fixture. The zirconia abutment with a titanium connection was subsequently introduced to maintain the titanium to titanium implant abutment interface. There are various designs of these abutments available in the market. The purpose of this study was to compare three types of zirconia abutments with different mechanisms of retention of the zirconia to the titanium interface. 15 Implants and abutments (3 Groups: Friction fit connection (Frft), Cemented (Cem) connection and titanium ring (Ti Ring) friction fit connection) were thermally cycled in water for 15,000 cycles and then cyclically loaded for 200,000 cycles or until failure at a frequency of 2 Hz using a sequentially increased loading protocol up to a maximum of 720 N. The mean load to failure values were as follows:The mean load to failure value for Group N was 526 N The mean load to failure value for Group S was 605 N The mean load to failure value for Group Z was 288 N. The abutment with the cemented connection showed the highest load to failure value and the abutment with the titanium ring friction fit connection showed the lowest load to failure value.The one-way analysis of variance was performed using the SAS procedure MIXED (SAS Institute Inc., Cary, NC, USA).There was a statistically significant difference between the three types of abutments tested (p<0.05).

Degree

thesis:*
Name thesis:degree_name
Master of Science
Level thesis:degree_level
masters
Discipline thesis:degree_discipline
Dentistry
Grantor dc:publisher
The Ohio State University
Year dc:date
2015

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Mascarenhas, Faye
Advisor dc:contributor.advisor
  • Yilmaz, Burak

Subjects

dc:subject × 1

Rights

dc:rights
Statement dc:rights
  • unrestricted
  • This thesis or dissertation is protected by copyright: all rights reserved. It may not be copied or redistributed beyond the terms of applicable copyright laws.
Language dc:language
English

Identifiers

dc:identifier.*

Chain of custody

source
Harvested from
OhioLINK
Base URL
etd.ohiolink.edu/acprod/odb_etd/ws/oai/oai
Last updated
2026-11-27
Source record
OAI-PMH GetRecord
related terms
citation

Mascarenhas, Faye. The Effect of Cyclic Loading on the Zirconia/Titanium Implant Abutment Interface and the Mechanism of Failure between Three Different Types of Abutments. masters thesis, The Ohio State University, 2015. http://rave.ohiolink.edu/etdc/view?acc_num=osu1440122618