Design and Testing of a Minimally Invasive Blood Clot Removal Device Constructed With Elements of Superelastic Nitinol

Abstract

dc:description

Many vascular system problems stem from insufficient blood return flow to the heart. One of the main causes is a blockage within veins or arteries known as a blood clot, or thrombus. This can occur after trauma, surgery, or other phenomenological reasons. Each year in the U.S. more than 175,000 bypass procedures and more than 160,000 amputations resulting from peripheral vessel disease are performed. Clinical data indicates that clot removal devices and procedures can reduce the need for an amputation by 80 percent. Percutaneous thrombectomy refers to the removal of thrombus using catheter based non-surgical methods. The ultimate goal of any modality to treat these conditions of the arterial or venous system is to restore patency, quickly, safely, and cost effectively. Catheter directed thrombectomy and thrombolysis is less traumatic and avoids the morbidity and mortality associated with conventional surgical technique. As a result, there has been a push recently for the use of percutaneous mechanical thrombectomy (PMT) devices. However, all devices have their own set of drawbacks: distal embolization, vessel wall trauma, hemolysis, to name a few. Ongoing efforts have been made to create a prototype thrombectomy device that uses elements of superelastic nitinol (a type of shape memory alloy), that seeks to address some of the drawbacks of current devices. The prototype was designed and tested in a simulated human circulatory system along side a commercially available device (The DiverCE Clot Extraction Catheter). The test evaluated how well the devices minimized distal embolization of a human blood clot created in vitro. Results of the testing showed that the prototype device created significantly less embolization when compared to the DiverCE particles greater than 102µm (p = 0.0332). Means were statistically not different for particles between 25µm and 102µm (p = 0.2454), and particles between 5µm and 25µm (p = 0.6524). In addition the prototype was shown to create insignificant embolization when compared to a control (p = 0.108). The DiverCE, on the other hand, was shown to create significant embolization when compared to a control (p = 0.027).

Degree

thesis:*
Name thesis:degree_name
Master of Science
Level thesis:degree_level
masters
Discipline thesis:degree_discipline
Mechanical Engineering
Grantor dc:publisher
University of Toledo
Year dc:date
2014

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Puffer, Andrew James
Chairs dc:contributor.committeechair
  • Elahinia, Mohammad
  • Cooper, Christopher

Subjects

dc:subject × 8

Rights

dc:rights
Statement dc:rights
  • unrestricted
  • This thesis or dissertation is protected by copyright: all rights reserved. It may not be copied or redistributed beyond the terms of applicable copyright laws.
Language dc:language
English

Identifiers

dc:identifier.*

Chain of custody

source
Harvested from
OhioLINK
Base URL
etd.ohiolink.edu/acprod/odb_etd/ws/oai/oai
Last updated
2026-11-27
Source record
OAI-PMH GetRecord
citation

Puffer, Andrew James. Design and Testing of a Minimally Invasive Blood Clot Removal Device Constructed With Elements of Superelastic Nitinol. masters thesis, University of Toledo, 2014. http://rave.ohiolink.edu/etdc/view?acc_num=toledo1399558753