University of Cincinnati
POLYMER FLIP-CHIP BONDING OF PRESSURE SENSORS ON FLEXIBLE KAPATON FILM FOR NEONATAL CATHETERS
Abstract
dc:descriptionThe object of this thesis is to develop a new packaging method to mount silicon micro pressure sensors into 1.67 mm diameter neonatal catheters to measure blood pressure. A new polymer flip-chip bonding method on a flexible Kapton film has been developed and applied to dual lumen neonatal catheters integrated with silicon micro pressure sensors. Flip chip bonding technique has inherent advantages of miniaturization, improved reliability and cost reduction. This polymer flip-chip bonding technique requires a low temperature process and improved tolerance of thermal stress which are very desirable for mounting micro sensors or micro actuators on lower-cost flexible polymer substrates for medical applications. The silicon micro pressure sensors, which are mounted on a flexible Kapton film with metal lines for neonatal catheter, have been fully characterized in both gas and liquid environments, indicating that the packaging was satisfactory rugged for catheter use.
Degree
thesis:*- Name thesis:degree_name
- MS
- Level thesis:degree_level
- masters
- Discipline thesis:degree_discipline
- Engineering : Electrical Engineering
- Grantor dc:publisher
- University of Cincinnati
- Year dc:date
- 2004
Author and committee
dc:creator, dc:contributor.*- Author dc:creator
-
- LI, CHUNYAN
- Advisor dc:contributor.advisor
-
- Ahn, Dr. Chong H.
Subjects
dc:subject × 3Rights
dc:rights- Statement dc:rights
-
- unrestricted
- This thesis or dissertation is protected by copyright: all rights reserved. It may not be copied or redistributed beyond the terms of applicable copyright laws.
- Language dc:language
- English
Identifiers
dc:identifier.*- Repository record dc:identifier
- http://rave.ohiolink.edu/etdc/view?acc_num=ucin1085764407