University of Cincinnati
A Comparison of the Accuracy of Measurement of Urine Thiocyanate by Spectrophotometry and Ion Chromatography in a Population Without Known Exposure to Cyanide
Abstract
dc:descriptionBackground: Urine thiocyanate(UTC) has been used as a marker of cyanide exposure. In 1996, 65 of 75(87%) workers in a TN plant had elevated UTC. After this, the testing laboratory(NMS) changed their testing method from spectrophotometry(SP) to ion chromatography(IC). Subsequently, only 1 of 13(8%) workers had elevated UTC. This prompted further investigation. Methods: 55 subjects without exposure to cyanide were tested for UTC by SP and IC. Known amounts of thiocyanate were added to a sample, and the resulting specimens were tested. A survey of laboratories was conducted. Results: The mean UTC level(µg/ml) by IC was 3.24 (σ = 3.75). The mean UTC level by SP was 16.84 (σ = 9.78). The RMS error was 8.06 for IC. The RMS error was 60.31 for SP. Three laboratories perform approximately 2400 UTC tests yearly in the US. Conclusions: The UTC testing by NMS in 1996 using the SP method overestimated levels.
Degree
thesis:*- Name thesis:degree_name
- MS
- Level thesis:degree_level
- masters
- Discipline thesis:degree_discipline
- Medicine : Occupational Medicine
- Grantor dc:publisher
- University of Cincinnati
- Year dc:date
- 2006
Author and committee
dc:creator, dc:contributor.*- Author dc:creator
-
- Case, Brian
- Contributors dc:contributor
-
- Freeman, Andrew
Rights
dc:rights- Statement dc:rights
-
- unrestricted
- This thesis or dissertation is protected by copyright: all rights reserved. It may not be copied or redistributed beyond the terms of applicable copyright laws.
- Language dc:language
- English
Identifiers
dc:identifier.*- Repository record dc:identifier
- http://rave.ohiolink.edu/etdc/view?acc_num=ucin1140464191