<i>In Vitro</i> Measurement and Calculation of Drag Force on Aortic Stentgraft in a Compliant Arterial Wall Model

Abstract

dc:description

Interventional treatment of aortic aneurysms using endovascular stent grafting is a minimally invasive technique. Following device implantation, continuous hemodynamic forces are exerted (drag force) on the stentgraft. When the drag force exceeds the fixation force, complications like stentgraft migration, endoleaks and stentgraft failure occur. In such a scenario the device becomes unstable, causing concern over the long-term durability of endovascular repairs. The objective of this study is: 1) to measure the drag force on a stentgraft in an in vitro experiment; 2) to calculate the drag force using blood flow–compliant arterial wall interaction model and compare it with the measured values on the stentgraft for the in vitro experiment; 3) to calculate drag force on the stentgraft using physiological flow conditions. Experimental data for a stentgraft within a silicon tubing, representing a compliant artery, shows a peak drag force of 2.79 N whereas the calculation predicts a peak drag force of 2.57 N; thus a percentage difference of 7.8 % is observed. When physiological flow and pressure pulse are used for the blood flow-compliant arterial wall computations, a peak drag force of 0.59 N is obtained for the same stentgraft that was used in the experiment. Compared to previous studies of constant iliac limb diameter for the abdominal aortic aneurysm stentgraft, the proposed model, where the distal end is not fixed to the arterial wall and is approximately half of its proximal diameter, the drag force is reduced by 47 % for the in vitro experiment and 81 % for the physiologic flow conditions. These forces can be used as design guidelines for the fixation force needed at the proximal end of the stentgraft under physiological pulsatile flow.

Degree

thesis:*
Name thesis:degree_name
MS
Level thesis:degree_level
masters
Discipline thesis:degree_discipline
Engineering : Mechanical Engineering
Grantor dc:publisher
University of Cincinnati
Year dc:date
2006

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Rontala Raghunathan, Ravi Shankar
Advisor dc:contributor.advisor
  • Banerjee, Dr. Rupak

Subjects

dc:subject × 3

Rights

dc:rights
Statement dc:rights
  • unrestricted
  • This thesis or dissertation is protected by copyright: all rights reserved. It may not be copied or redistributed beyond the terms of applicable copyright laws.
Language dc:language
English

Identifiers

dc:identifier.*

Chain of custody

source
Harvested from
OhioLINK
Base URL
etd.ohiolink.edu/acprod/odb_etd/ws/oai/oai
Last updated
2026-11-27
Source record
OAI-PMH GetRecord
citation

Rontala Raghunathan, Ravi Shankar. <i>In Vitro</i> Measurement and Calculation of Drag Force on Aortic Stentgraft in a Compliant Arterial Wall Model. masters thesis, University of Cincinnati, 2006. http://rave.ohiolink.edu/etdc/view?acc_num=ucin1163703097