Back to search

University of Cincinnati

The Validation of LeadCare © II Portable Blood Lead Analyzer through Graphite Furnace Absorption Spectroscopy (GFAAS)

Abstract

dc:description

This study correlates a LeadCare © II blood analyzer with Graphite Furnace Absorption Spectroscopy (GFAAS) and performs a comparison of the lead level results. Subjects who currently occupied a lead containing building that was under construction were recruited. The study consisted of 12 total participants. One sample was collected from each subject via venipuncture. The samples were then analyzed with both a LeadCare © II analyzer and also via Graphite Furnace Absorption Spectroscopy (GFAAS). The sample mean for the GFAAS analysis was 1.56 µg/dL with a SD ± 1.93. Many of the samples were reported at or below the limit of detection of 1.0 µg/dL. The reportable range of the Lead © II device is 3.3 to 65 µg/dL with most of the results being reported <3.3 µg/dL. Due to the analysis methods having different reportable ranges and most being below these measurable spectrum, it was considered illogical to make a comparison of measurements between these two devices.

Degree

thesis:*
Name thesis:degree_name
MS
Level thesis:degree_level
masters
Discipline thesis:degree_discipline
Medicine: Industrial Hygiene (Environmental Health)
Grantor dc:publisher
University of Cincinnati
Year dc:date
2015

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Fries, Michael N.
Chair dc:contributor.committeechair
  • Talaska, Glenn

Subjects

dc:subject × 1

Rights

dc:rights
Statement dc:rights
  • unrestricted
  • This thesis or dissertation is protected by copyright: some rights reserved. It is licensed for use under a Creative Commons license. Specific terms and permissions are available from this document's record in the OhioLINK ETD Center.
Language dc:language
English

Identifiers

dc:identifier.*

Chain of custody

source
Harvested from
OhioLINK
Base URL
etd.ohiolink.edu/acprod/odb_etd/ws/oai/oai
Last updated
2026-11-27
Source record
OAI-PMH GetRecord
citation

Fries, Michael N.. The Validation of LeadCare © II Portable Blood Lead Analyzer through Graphite Furnace Absorption Spectroscopy (GFAAS). masters thesis, University of Cincinnati, 2015. http://rave.ohiolink.edu/etdc/view?acc_num=ucin1439305875