University of Cincinnati
Efficacy of New Diagnostic Parameters for Determining Arteriovenous Fistula Functionality: A Numerical Study
Abstract
dc:descriptionIntroduction. The current diagnostic endpoints that are in use to assess the functionality of Arteriovenous Fistula (AVF) are based on either flow or pressure. These parameters, namely vascular access blood flow rate (Qa) and ratio of venous pressure to arterial pressure (VAPR) have been found to provide an inaccurate estimate of AVF functionality in the presence of a venous stenosis. These indices cannot predict the patency of fistula in the future. Recently, two new diagnostic parameters, resistance index (R: ratio of pressure drop across the AVF to flow rate) and pressure drop coefficient (Cp: ratio of pressure drop across the AVF to dynamic pressure) that combine the effect of both pressure and flow, were tested in porcine models. In this study, the effectiveness of these two parameters was studied in a controlled hemodynamic environment on an idealized AVF geometry for different percentage area stenosis (%AS: 0%, 65%, 75%, and 85%) and time-averaged flow rate at proximal artery (Q ¯pa: 600, 1000, and 1600 mL/min). Methods. An idealized AVF geometry was developed using the geometrical data from the AVFs that were created in the porcine models. Computational fluid dynamics (CFD) analyses were performed on the idealized AVF geometry for each of the four stenosis cases and the three Q ¯pa. Both R and Cp were calculated for the twelve cases (four %AS and three Q ¯pa). Results. It was observed that both R and Cp elevated with the increase in %AS for a fixed Q ¯pa. For a fixed AS, the R decreased with the reduction in Q ¯pa, while Cp increased.Conclusion. Reduction of Q ¯pa and an increase in %AS are indicative of a deteriorating AVF functionality, also known as adverse remodeling. The decrease in the value of R with the reduction in Q ¯pa for a fixed %AS can be deduced as a drawback of this parameter. On the other hand, the increase in Cp with the increase in %AS (when Q ¯pa is fixed) and the decrease in Q ¯pa (%AS is fixed) indicates that Cp can detect adverse remodeling in AVFs. Thus, Cp can provide a better prediction of adverse remodeling in AVFs as a result of an increasing stenosis, reducing flow rate, or both.
Degree
thesis:*- Name thesis:degree_name
- MS
- Level thesis:degree_level
- masters
- Discipline thesis:degree_discipline
- Engineering and Applied Science: Mechanical Engineering
- Grantor dc:publisher
- University of Cincinnati
- Year dc:date
- 2016
Author and committee
dc:creator, dc:contributor.*- Author dc:creator
-
- Subramony Anantha, Krishna
- Chair dc:contributor.committeechair
-
- Banerjee, Rupak
Subjects
dc:subject × 8Rights
dc:rights- Statement dc:rights
-
- unrestricted
- This thesis or dissertation is protected by copyright: some rights reserved. It is licensed for use under a Creative Commons license. Specific terms and permissions are available from this document's record in the OhioLINK ETD Center.
- Language dc:language
- English
Identifiers
dc:identifier.*- Repository record dc:identifier
- http://rave.ohiolink.edu/etdc/view?acc_num=ucin1470753788