Determination Of Dose Effects When Including Attenuation Of The Treatment Table Into Treatment Planning Computer Modeling

Abstract

dc:description

During radiation treatment, the patient commonly lies on a horizontal treatment table called a couch. The material used for the couch is chosen during the manufacturing process so as to minimize the effect of the couch on dose delivery, but it still has an effect. In addition to couch attenuation on the beam, the modern approach of using multiple beam angles for dose delivery makes treatment planning in the presence of the couch more challenging. However, the effect of the couch is not always considered when creating treatment plans, generally due to treatment planning software limitations. In this evaluation, the effects of the Varian Medical Systems, Inc® (Palo Alto, CA) Model IGRT Couch Top® on the dose distribution of head and neck cancer treatment was determined for a linear accelerator operating at 6 MV. Four head and neck cancer patients were selected for the study. Varian Model Eclipse® treatment planning software was used to create two Volumetric Modulated Arc Therapy (VMAT) plans. One plan assumed the couch was not present, and one assumed it was present. Both plans were optimized to achieve the same physician directed dose objectives. Comparisons were then made using the planning system’s Dose Volume Histogram (DVH) statistical and volumetric data. Dose changes to the tumor volumes along with normal tissue structures around the target area were clinically assessed. The mean dose to the tumor was reduced by 0.5 to 0.65 Gy as a consequence of including the couch during planning. Doses to surrounding normal tissues showed variable, yet significant effects.

Degree

thesis:*
Name thesis:degree_name
Master of Science (MS)
Level thesis:degree_level
masters
Discipline thesis:degree_discipline
Physics
Grantor dc:publisher
Wright State University
Year dc:date
2018

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Alshamrany, Abdullah
Contributors dc:contributor
  • Foy, Brent

Subjects

dc:subject × 4

Rights

dc:rights
Statement dc:rights
  • unrestricted
  • This thesis or dissertation is protected by copyright: some rights reserved. It is licensed for use under a Creative Commons license. Specific terms and permissions are available from this document's record in the OhioLINK ETD Center.
Language dc:language
English

Identifiers

dc:identifier.*

Chain of custody

source
Harvested from
OhioLINK
Base URL
etd.ohiolink.edu/acprod/odb_etd/ws/oai/oai
Last updated
2026-11-26
Source record
OAI-PMH GetRecord
citation

Alshamrany, Abdullah. Determination Of Dose Effects When Including Attenuation Of The Treatment Table Into Treatment Planning Computer Modeling. masters thesis, Wright State University, 2018. http://rave.ohiolink.edu/etdc/view?acc_num=wright1520859358953399