Back to search

Oklahoma State University

Measurements of Photo-neutrons from a Medical Linear Accelerator Using Cr-39 Plastic Nuclear Track Dtetectors

Abstract

dc:description.abstract

Photo-neutrons are produced when x-ray energies exceed 7 MeV. Photo-neutron production varies depending on x-ray beam energy. CR-39 PNTDs were used in this study to measure the neutron absorbed dose and dose equivalent produced by a Varian Clinac 23EX for x-ray beams of 6 and 18 MVp and with a Varian Trilogy using an x-ray beam of 10 MVp. Neutron absorbed dose and dose equivalent were measured at 100 cm SSD at 0, 20, and 40 cm off-axis from the primary beam in air. Using a polyethylene phantom the neutron absorbed dose and dose equivalent were measured at 100 cm SSD from the top of the phantom at 0, 5, and 10 cm from the surface, in the beam central axis and off-axis distances of 20 and 40 cm at a depth of 10 cm. The neutron absorbed dose and dose equivalent from medical linear accelerators have been measured from the LET spectrum of recoiled tracks produced in the CR-39 PNTDs for high energy neutrons (1-20 MeV) and the neutron dose equivalent for low energy (< 1 MeV) neutrons were measured using TRNDs. In this study, the experimentally measured absorbed dose and dose equivalent from photo-neutrons produced in a common medical linear accelerator operating at energies 10 and 18 MVp range from 0.2 �Gyn/Gyx and 2 �Svn/Gyx to 495 �Gyn/Gyx and 5500 �Svn/Gyx respectively. The neutron absorbed dose for the 6 MVp x-ray beam was measured to be 0 �Gyn/Gyx. Low energy neutrons accounted for less than 1% of the neutron dose equivalent. Those detectors exposed inside of the phantom measured a higher contribution from high LET (> 100 keV/�m) particles than those detectors exposed in air.

Degree

thesis:*
Department dc:contributor.department
Physics
Grantor dc:publisher
Oklahoma State University
Year dc:date.issued
2014

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Monson, Jonathan Michael
Advisor dc:contributor.advisor
  • Benton, Eric R.
Committee members dc:contributor.committeemember
  • Yukihara, Eduardo
  • Peakheart, David

Rights

dc:rights.accessrights
Statement dc:rights.accessrights
  • Open Access
  • Copyright is held by the author who has granted the Oklahoma State University Library the non-exclusive right to share this material in its institutional repository. Contact Digital Library Services at lib-dls@okstate.edu or 405-744-9161 for the permission policy on the use, reproduction or distribution of this material.
Language dc:language
en_US

Identifiers

dc:identifier.*
Mods Identifier Filename
Monson_okstate_0664M_13253.pdf

Chain of custody

source
Harvested from
Oklahoma State University
Base URL
openresearch.okstate.edu/server/oai/request
Last updated
2026-11-26
Source record
OAI-PMH GetRecord
related terms
citation

Monson, Jonathan Michael. Measurements of Photo-neutrons from a Medical Linear Accelerator Using Cr-39 Plastic Nuclear Track Dtetectors. Oklahoma State University, 2014. https://hdl.handle.net/20.500.14446/15023