Back to search

Oklahoma State University

Design and Testing of Collimators for Use in Measuring X-ray Attenuation

Abstract

dc:description.abstract

The main goal of this project was to design and test a set of collimators for use in measuring x-ray attenuation. Two collimators: a primary collimator and a secondary collimator, were built. The primary collimator was built using 10.0 cm of lead, followed by 7.0 cm of tin, 1.5 mm of copper and 0.5 mm of aluminum, in the respective order. A 1.5 mm thick, copper filter was also used on the aperture of the primary collimator. The secondary collimator was built using 2.5 cm of lead. The cross-sectional size of both the collimators was made to be 15 x 15 cm2. The shape of the aperture in each collimator was made to be 1.5 cm x 3.0 cm. Various features of the collimators were tested in the Varian TrueBeam and Varian Trilogy linear accelerators. The primary collimator was found to best minimize the secondaries produced by the various components inside the gantry and the collimator itself that reach the absorber when placed at least 30.0 cm away from the absorber upstream the beam. The copper filter in the primary collimator was found unnecessary. The secondary collimator was found to best minimize the secondaries produced by the absorber that reach the detector, when placedat least 20.0 cm away from the detector upstream the beam. With both the collimators placed at their optimal positions, the cross-sectional size of the collimated beam measured around the detector, placed at 200 cm from the source, was found to be approximately equal to 1.5 cm x 3.0 cm. The cross-sectional size of the collimated beam at the position of absorber was found to be 1.9 cm x 4.3 cm and that of attenuated x-ray beam was found to be 3.0 x 5.5 cm.

Degree

thesis:*
Department dc:contributor.department
Physics
Year dc:date.issued
2017

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Panthi, Rajesh
Advisor dc:contributor.advisor
  • Benton, Eric R.
Committee members dc:contributor.committeemember
  • Yukihara, Eduardo G.
  • Borunda, Mario F.

Rights

dc:rights.accessrights
Statement dc:rights.accessrights
  • Open Access
  • Copyright is held by the author who has granted the Oklahoma State University Library the non-exclusive right to share this material in its institutional repository. Contact Digital Library Services at lib-dls@okstate.edu or 405-744-9161 for the permission policy on the use, reproduction or distribution of this material.
Language dc:language
en_US

Identifiers

dc:identifier.*
Mods Identifier Filename
PANTHI_okstate_0664M_15531.pdf

Chain of custody

source
Harvested from
Oklahoma State University
Base URL
openresearch.okstate.edu/server/oai/request
Last updated
2026-11-26
Source record
OAI-PMH GetRecord
related terms
citation

Panthi, Rajesh. Design and Testing of Collimators for Use in Measuring X-ray Attenuation. 2017. https://hdl.handle.net/20.500.14446/300095