Abstract

dc:description.abstract

The prompt gamma-ray timing (PGT) method offers a relatively simple approach for range verification in proton therapy. Starting from the findings of previous experiments, several steps toward a clinical application of PGT have been performed in this work. First of all, several scintillation materials have been investigated in the context of PGT. The time resolution was determined at high photon energies in the MeV-region. In conclusion, the fast and bright scintillator CeBr3 is the material of choice in combination with a timing photomultiplier tube as light detector. A second study was conducted at Universitäts Protonen Therapie Dresden (UPTD) to characterize the proton bunch structure of a clinical beam concerning its time width and relative arrival time. The data is mandatory as input for simulation studies and to correct for phase drifts. The obtained data could furthermore be used for the first 2D imaging of a heterogeneous phantom based on prompt gamma-rays. In a last step, a PGT prototype system was designed using the findings from the first two studies. The prototype system is based on a newly developed digital spectrometer and a CeBr3 detector. The device is characterized at the ELBE bremsstrahlung beam. It was verified that the prototype operates within the specifications concerning time and resolution as well as throughput rate. Finally, for the first time the PGT system was used under clinical conditions in the treatment room of UPTD. Here, PGT data was obtained from the delivery of a three-dimensional treatment plan onto PMMA phantoms. The spot-by-spot analysis helped to investigate the performance of the prototype device under clinical conditions. As a result, range variations of 5 mm could be detected for the first time with an uncollimated system at clinically relevant doses. To summarize, the obtained results help to bring PGT closer to a clinical application.

Degree

thesis:*
Level thesis:degree_level
thesis.doctoral
Department thesis:degree_department
Fakultät Mathematik und Naturwissenschaften
Grantor dc:publisher
Technische Universität Dresden
Year dc:date.issued
2018

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Petzoldt, Johannes
Advisor dc:contributor.advisor
  • Zuber, Kai

Subjects

dc:subject × 5

Rights

dc:rights.accessrights
Statement dc:rights.accessrights
  • free
  • info:eu-repo/semantics/openAccess
Language dc:language.dcterms_iso639_2
eng

Identifiers

dc:identifier.*
Dc Publisher Identifier
4519974-7
Ddb Identifier Erstkat Id
1013213629

Chain of custody

source
Harvested from
QUCOSA
Base URL
www.qucosa.de/oai/
Last updated
2026-11-27
Source record
OAI-PMH GetRecord
citation

Petzoldt, Johannes. Toward the Clinical Application of the Prompt Gamma-Ray Timing Method for Range Verification in Proton Therapy. thesis.doctoral thesis, Technische Universität Dresden, 2018. https://nbn-resolving.org/urn:nbn:de:bsz:14-qucosa-231987