Technische Universität Dresden
Application of experimental and analytical approaches in characterizing coronary stents
Abstract
dc:description.abstractCoronary artery disease (CAD) affects every fifth person in the world. The gold-standard treatment for CAD is stent implantation, however, the existing therapy is not sufficient due to many reasons. For instance, in-stent restenosis, biocompatibility, controlled degradation rate, protein adsorption, and adequate endothelialization are still the main concerns. In the last two decades, the field of stent technology has been grown rapidly and many new stent types and in vitro testing methods for stent characterization have been developed to minimize the aforementioned issues. In this vicinity, there are still many unaddressed issues: i) the quantitative analysis of corrosion is conducted with simpler samples made of stent material instead of stents, in most cases due to the absence of a mathematical model to calculate the entire stent surface area (ESSA); ii) in vitro stent testing in environments that are very far from actual physiological environments; iii) Evaluation of the influence of in-vitro test conditions on coated metallic stents; iv) absence of flow-induced shear stress (FISS) corrosion model, to mention a few. This thesis presents the novel ESSA model, the fluid dynamic experimental setup with the integration of various sensors and pH control, the influence of in vitro degradation behavior of the titanium oxynitride (TiOxNy) coated stainless steel stents and anodized AZ31 samples, and the FISS corrosion model. The results show some important contributions in this field, however, there is still a huge potential for the development of promising stent characterization solutions.
Degree
thesis:*- Level thesis:degree_level
- thesis.doctoral
- Department thesis:degree_department
- Maschinenwesen
- Grantor dc:publisher
- Technische Universität Dresden
- Year dc:date.issued
- 2023
Author and committee
dc:creator, dc:contributor.*- Author dc:creator
-
- Saqib, Muhammad
Subjects
dc:subject × 25- stent
- corrosion
- coronary artery disease
- fluid dynamic
- flow induced shear stress corrosion
- entire stent surface area
- magnetron sputtering
- bare metal stent
- metallic stent
- in vitro testing
- mg alloy
- titanium oxynitride
- coated stents
- Korrosion
- koronare Herzkrankheit
- Fluiddynamik
- strömungsinduzierte Scherspannungskorrosion
- gesamte Stentoberfläche
- Magnetronsputtern
- blanker Metallstent
- metallischer Stent
- In-vitro-Tests
- mg-Legierung
- Titan-Oxynitrid
- beschichtete Stents
Rights
dc:rights.accessrights- Statement dc:rights.accessrights
-
- free
- info:eu-repo/semantics/openAccess
- Language dc:language.dcterms_iso639_2
- eng
Identifiers
dc:identifier.*- Dc Publisher Identifier
- 2024114-8
- Ddb Identifier Erstkat Id
- 1852092122