Clinical implementation of quality-assured and optimized pencil beam scanning proton therapy for moving targets

Abstract

dc:description.abstract

Radiotherapy is one of the major treatment methods for cancer patients. Besides conventional radiotherapy with photons, proton therapy allows for a more conformal dose distribution with higher sparing of normal tissue. However, the application of proton therapy for targets undergoing respiratory motion is limited due to uncertainties in the delivered dose caused by motion and interplay effects. The assessment of the delivered dose distributions including these motion effects as well as treatment planning with sufficient robustness against respiratory motion are prerequisites for high quality proton therapy for moving targets. Within this thesis, an algorithm for quality assurance was developed which reconstructs the delivered dose distribution considering the respiratory motion that occurred during the dynamic beam delivery. The algorithm was experimentally validated and provided accurate results under realistic clinical conditions for peak-to-peak motion amplitudes of up to 15 mm. Following the clinical implementation, the first patients receiving a fraction-wise dose reconstruction were analysed in regards to intra- and interfractional effects. While intrafractional dose deviations were averaged out within a few fractions, interfractional effects led to a serious loss in target coverage in some cases. The process of the dose reconstruction took approximately 15-20 minutes per fraction and is thus feasible for routine application. To determine the best treatment planning strategy for moving targets, five different robust optimization methods were investigated for a cohort of lung cancer patients with considerable target motion. The plan robustness against setup and range uncertainties, intra- and interfractional changes as well as interplay effects was compared. All plans offered a similar plan robustness and none of the investigated strategies could maintain sufficient target coverage in the interplay-affected dose distributions for target motion amplitudes of more than 10 mm. Most benefits were found for the four dimensional (4D) robust optimization based on the average computed tomography (CT) and three 4DCT phases. When evaluating plan robustness against respiratory motion, the method which neglected interplay effects largely overestimated the target coverage.

Degree

thesis:*
Level thesis:degree_level
thesis.doctoral
Department thesis:degree_department
Medizinische Fakultät Carl Gustav Carus
Grantor dc:publisher
Technische Universität Dresden
Year dc:date.issued
2025

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Spautz, Saskia

Subjects

dc:subject × 10

Rights

dc:rights.accessrights
Statement dc:rights.accessrights
  • free
  • info:eu-repo/semantics/openAccess
Language dc:language.dcterms_iso639_2
eng, ger

Identifiers

dc:identifier.*
Dc Publisher Identifier
2024114-8

Chain of custody

source
Harvested from
QUCOSA
Base URL
www.qucosa.de/oai/
Last updated
2026-11-27
Source record
OAI-PMH GetRecord
citation

Spautz, Saskia. Clinical implementation of quality-assured and optimized pencil beam scanning proton therapy for moving targets. thesis.doctoral thesis, Technische Universität Dresden, 2025. https://nbn-resolving.org/urn:nbn:de:bsz:14-qucosa2-963828