Back to search

Queen's University at Kingston

Rotation Encoding of C-Arm Fluoroscopes With Accelerometer

Abstract

dc:description.abstract

Accurate, practical, and affordable pose tracking on manually operated C-arm fluoroscopes is a major technical challenge. Conventional tracking methods, such as optical cameras and radiographic fiducials, are hampered by significant shortcomings. Optical cameras are delicate, costly, and have a complex system setup that is easily susceptible to camera obstruction in cluttered operating room. Radiographic fiducials occupy a significant portion of the fluoroscopic imaging space. Using fiducials also requires segmentation that limits clinical use. In this thesis, an alternative form of tracking is proposed to encode the rotational joints of manually operated C-arms using a tilt sensing accelerometer for tracking the C-arm rotational pose. The technique is evaluated by affixing an accelerometer to a full-scale C-arm where a webcam is used as a substitute for X-ray imaging. Ground truth C-arm rotational poses were obtained from the webcam by tracking a checkerboard plate. From these rotational poses, a series of angle and structural correction equations were formulated that can properly relate the accelerometer angle readings to the C-arm rotational pose in real-time and compensate for systematic structural C-arm deformations, such as sagging and bending. Real-time rotational pose tracking of the primary and secondary joint rotations of the C-arm showed an accuracy of 0.5 degrees in the entire range of interest.

Degree

thesis:*
Department dc:contributor.department
Electrical and Computer Engineering
Grantor
Queen's University at Kingston
Year dc:date.issued
2011

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Grzeda, Victor
Advisor dc:contributor.supervisor
  • Fichtinger, Gabor

Subjects

dc:subject × 5

Rights

Language dc:language.iso
eng

Identifiers

dc:identifier.*
Handle dc:identifier.uri
https://hdl.handle.net/1974/6288

Chain of custody

source
Harvested from
Queens University
Base URL
qspace.library.queensu.ca/server/oai/request
Last updated
2026-11-19
Source record
OAI-PMH GetRecord
citation

Grzeda, Victor. Rotation Encoding of C-Arm Fluoroscopes With Accelerometer. Queen's University at Kingston, 2011. https://hdl.handle.net/1974/6288