Calibration of a Mobile-Gantry CT Scanner for Surgical Navigation
Abstract
dc:description.abstractIn image-guided surgical navigation, instruments tools are tracked by a position sensor and their locations rendered along with the patient's anatomy. Conventional methods require an invasive, time-consuming and potentially uncertain process of intra-operative registration of the images to the patient. In a direct navigation system, such an intra-operative registration is replaced with pre-operative patient-independent calibration in a process that determines the relationship between the coordinate frame of imaging equipment and the coordinate frame of the position sensor. This dissertation presents a method for pre-operatively calibrating a direct navigation system that used an optical position sensor and a mobile gantry CT scanner. A custom bi-local calibration device was designed and manufactured, after which existing navigation software was augmented with components that used the pre-operatively determined transformation to provide image-guided surgical navigation. The resulting system was tested in an image-guided operating suite using plastic bone models. In the validation stage, the inherent error was less than 0.4 mm and the target registration error was approximately 1.6 mm for a ceiling-mounted position sensor and 0.7 mm for a portable position sensor. This accuracy is consistent with the best intra-operative registrations reported in the literature and this calibration method may be useful in future surgical systems.
Degree
thesis:*- Department dc:contributor.department
- Computing
- Grantor
- Queen's University at Kingston
- Year dc:date.issued
- 2013
Author and committee
dc:creator, dc:contributor.*- Author dc:creator
-
- Belkova, Anna
- Advisor dc:contributor.supervisor
-
- Ellis, Randy E.
Subjects
dc:subject × 3Rights
- Language dc:language.iso
- eng
Identifiers
dc:identifier.*- Handle dc:identifier.uri
- http://hdl.handle.net/1974/8252