Rice University
An Ultrasonic Transceiver for Non-Invasive Intracranial Pressure Sensing
Abstract
dc:description.abstractThis work presents a 9-mW ultrasonic through-transmission transceiver (TRX) designed for portable, non-invasive intracranial pressure (ICP) sensing. The system employs two ultrasound transducers placed at the temporal bone windows to measure changes in ultrasonic time-of-flight (ToF), from which skull expansion and the corresponding ICP waveform are derived. Key components include a high-efficiency Class-DE power amplifier (PA) with 95% efficiency and a 15.8 VPP output swing, as well as a successive approximation register (SAR) delay-locked loop (DLL)-based time-to-digital converter (TDC) featuring 29.8 ps resolution and a 122 ns range. In addition to electrical characterization, the sensor is validated through two demonstrations: a water tank setup and a human head phantom setup. The system shows a high correlation (R² = 0.93) with a medical-grade invasive ICP sensor. The proposed system offers high accuracy, low power consumption, and reliable performance, making it a promising solution for real-time, portable, non-invasive ICP monitoring in various clinical settings.
Degree
thesis:*- Name thesis:degree_name
- Master of Science
- Level thesis:degree_level
- Masters
- Discipline thesis:degree_discipline
- Engineering
- Department thesis:degree_department
- Electrical and Computer Engineering
- Grantor
- Rice University
- Year dc:date.issued
- 2025
Author and committee
dc:creator, dc:contributor.*- Author dc:creator
-
- Topalli, Gerald
- Advisor dc:contributor.advisor
-
- Chi, Taiyun
Subjects
dc:subject × 2Rights
dc:rights- Statement dc:rights
-
- Copyright is held by the author, unless otherwise indicated. Permission to reuse, publish, or reproduce the work beyond the bounds of fair use or other exemptions to copyright law must be obtained from the copyright holder.
- Language dc:language.iso
- eng
Identifiers
dc:identifier.*- Handle dc:identifier.uri
- https://hdl.handle.net/1911/118366