Calibration of a leadglass lead-scintillator photon detector at Fermilab experiment E-704 (Lead-scintillator)
Abstract
dc:description.abstractWe calibrated a leadglass Pb-scintillator photon detector with a positron beam of 30 GeV. The detector consisted of a segmented leadglass counter in front and a Pb-scintillator sandwich counter in the back. The response of the sandwich counter exhibited an exponential dependence on the position of the incident particle. The calibration constants of the leadglass segments were obtained by an iterative method. The results of our analysis showed an energy resolution of σ ({\rm E})\over {\rm E} = 24%/$\surd$E. A monitoring system using LED and $\sp{241}$Am sources regularly checked the stability of all modules. A typical photomultiplier's gain factor deviated no more than 6% from the mean value at any time during a period of three months.
Degree
thesis:*- Name thesis:degree_name
- Master of Science
- Level thesis:degree_level
- Masters
- Discipline thesis:degree_discipline
- Natural Sciences
- Department thesis:degree_department
- Physics
- Grantor
- Rice University
- Year dc:date.issued
- 1988
Author and committee
dc:creator, dc:contributor.*- Author dc:creator
-
- Nguyen, Chau T.
- Advisor dc:contributor.advisor
-
- Roberts, Jabus B.
Subjects
dc:subject × 3Rights
dc:rights- Statement dc:rights
-
- Copyright is held by the author, unless otherwise indicated. Permission to reuse, publish, or reproduce the work beyond the bounds of fair use or other exemptions to copyright law must be obtained from the copyright holder.
- Language dc:language.iso
- eng
Identifiers
dc:identifier.*- Dc Identifier Callnumber
- Thesis Phys. 1989 Nguyen
- Dc Identifier Citation
- Nguyen, Chau T.. "Calibration of a leadglass lead-scintillator photon detector at Fermilab experiment E-704 (Lead-scintillator)." (1988) Master’s Thesis, Rice University. <a href="https://hdl.handle.net/1911/13383">https://hdl.handle.net/1911/13383</a>.