Rice University
Calibration of the STAR Time-of-Flight Detector for Particle Identification
Abstract
dc:description.abstractThis thesis describes several calibrations of the Time-of-Flight (TOF) detector at the Solenoidal Tracker at RHIC (STAR). These calibrations are required to allow Particle Identification (PID) which benefits many physics analysis. These calibrations treat the Integral Non-Linearity (INL) of the time to digital converters, the global offsets, the slewing from pulse time dependence on the pulse height, and the transmission times inside the detectors. Each of these corrections will be described. The data for the INL correction was collected at Rice University using two different electronics configurations. Care was taken to insure the two approaches were consistent. These calibrations were tested using "cable-delay tests." Data from RHIC Run 9 200 GeV data will be used to confirm the effectiveness of the INL, offset, slewing, and transmission time calibrations.
Degree
thesis:*- Name thesis:degree_name
- Master of Science
- Level thesis:degree_level
- Masters
- Discipline thesis:degree_discipline
- Natural Sciences
- Department thesis:degree_department
- Physics
- Grantor
- Rice University
- Year dc:date.issued
- 2011
Author and committee
dc:creator, dc:contributor.*- Author dc:creator
-
- McDonald, Daniel
- Advisor dc:contributor.advisor
-
- Llope, William J.
Subjects
dc:subject × 2Rights
dc:rights- Statement dc:rights
-
- Copyright is held by the author, unless otherwise indicated. Permission to reuse, publish, or reproduce the work beyond the bounds of fair use or other exemptions to copyright law must be obtained from the copyright holder.
- Language dc:language.iso
- eng
Identifiers
dc:identifier.*- Dc Identifier Citation
- McDonald, Daniel. "Calibration of the STAR Time-of-Flight Detector for Particle Identification." (2011) Master’s Thesis, Rice University. <a href="https://hdl.handle.net/1911/70345">https://hdl.handle.net/1911/70345</a>.
- Dc Identifier Other
- McDonaldD