Back to search

Rochester Institute of Technology

LVAD Occlusion Condition Monitoring Using State Augmented Acoustic Spectral Images

Abstract

dc:description.abstract

Each year, thousands of people die from heart disease and related illnesses due to the lack of available donor organs. Left ventricular assist devices (LVADs) aim to mitigate that occurrence, serving as a bridge-to-surgery option. While short term survival rates of LVAD patients near that of orthotopic surgery they are not viable long term options due to varied reasons. This work examines one cause, outlet graft thrombosis, and develops an algorithm for increasingly robust classification of device condition as it pertains to thrombosis or more generally occlusion. In order to do so an in vitro heart simulator is developed so that varying degrees of signal non-stationarity can be simulated and tested over a wide range of physiological blood pressure and heart rate conditions. Using a seeded-fault methodology, acoustics are acquired at the LVAD outlet graft location and subsequent spectral images of the sounds are developed. Statistical parameters from the images are used as features for classification using a support vector machine (SVM) which yields promising results. Given a comprehensive training space classification can be performed to fair accuracies (roughly 80%) using only the spectral image parameters. However, when the training space is limited augmenting the image features with patient state parameters elicits more robust identification. The algorithm developed in this work offers non-invasive diagnostic potential for LVAD conditions otherwise requiring invasive means.

Degree

thesis:*
Name thesis:degree_name
Mechanical Engineering (MS)
Level thesis:degree_level
Thesis
Discipline thesis:degree_discipline
Mechanical Engineering (KGCOE)

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Prechtl, Ian Richard
Contributors dc:contributor
  • Jason Kolodziej

Subjects

dc:subject × 6

Identifiers

dc:identifier.*
Repository record dc:identifier
https://repository.rit.edu/theses/9884

Chain of custody

source
Harvested from
Rochester Institute of Technology
Base URL
repository.rit.edu/do/oai/
Last updated
2026-11-19
Source record
OAI-PMH GetRecord
citation

Prechtl, Ian Richard. LVAD Occlusion Condition Monitoring Using State Augmented Acoustic Spectral Images. Thesis thesis, https://repository.rit.edu/theses/9884