Upper limb prosthetic socket-limb coupling: Novel measurement approaches and impacts on socket performance and comfort
Abstract
dc:description.abstractPeople with upper limb difference consistently report dissatisfaction with prosthetic comfort and performance, with these issues cited as key contributors to device abandonment or non-use. Advances in automating prosthetic manufacture and emerging interest in adjustable prosthetic sockets offer the potential to deliver better fitting sockets but design methods remain limited. Design approaches are limited by current understanding of how to define and evaluate socket fit, which in turn makes it difficult to explore relationships between socket fit (mechanical function), comfort, and performance. Socket mechanical function is the extent to which socket pose is maintained relative to the residuum under the range of mechanical disturbances seen in normal use. Previous attempts to measure socket mechanical function in the upper-limb have focused on trans-humeral different individuals, or intact-limb individuals, and none have evaluated the impact of varying socket fit on socket mechanical function. Varying socket fit is imperative in enabling the impact of fit on socket mechanical function, comfort, and performance, to be investigated, and through this better understand how ideal socket fit can be achieved through socket design.The aim of this thesis is to develop and validate a marker-based motion capture approach for investigating socket mechanical function, and to utilise this approach to better understand the effect varying socket fit has on socket comfort and performance. Measuring socket mechanical function requires the socket and forearm to be tracked independently, however the International Society of Biomechanics recommended approach for tracking the ulna requires wrist styloid markers and hence can’t be used in studies involving trans-radial limb different individuals. Study one investigated potential alternative ulna tracking frames in ten intact limb participants, and provided recommendations for 18 alternative frames, the optimal of which can be selected depending on residuum length.To support the experimental work in studies two and three, an adjustable, open-frame bypass socket was then designed, consisting of two fixed pads and one floating pad, connected by a cord, the effective length of which could be adjusted using a micro-adjustment dial. A spring, in-line with the cord, provided a measure of cord tension, which is used to quantify socket tightness.Study two reported on a new marker-based motion capture method for measuring socket mechanical function. The study systematically varied the tightness of the adjustable bypass socket and tracked the relative socket-limb movement under a range of time-varying loads applied along, and around the long axis of the socket. Four metrics of socket mechanical function were calculated: the maximum displacement under a given load; the socket-limb slip under load; the average stiffness of the socket-limb interface; and the hysteresis over a loading cycle. Ten intact-limb participants were recruited and the relationships between socket tightness and each of the four socket mechanical function metrics were analysed. Average stiffness was shown to have the strongest statistical relationship with socket tightness. The study also showed that the socket-limb interface demonstrated both increased stiffness with increased applied load, and hysteresis over repeated loading cycles, indicating viscoelastic behaviour. Such behaviours have previously been reported in indentation testing of forearm tissue, but not in the socket-limb interface.Study three investigated the practicality of measuring socket mechanical function, comfort, and performance within a single experimental protocol. Feasibility was demonstrated in three intact-limb participants.Finally, given the inherent limitations of surface-based marker methods, a preliminary study demonstrated the feasibility of radiographic measurement of socket-limb mechanical function as an alternative approach. In summary, this thesis establishes novel methods for quantifying upper-limb socket mechanical function under dynamic loading conditions. These methods provide a foundation for future investigations of the relationships between socket mechanical function, comfort, and performance, in transradial limb different individuals. Whilst the experimental work was limited to intact-limb participants, the methods increased understanding of the mechanical behaviour of the socket-limb interface.
Author and committee
dc:creator, dc:contributor.*- Author dc:creator
-
- Pickard, Nicolaas
- Author Identifier dc:creator.authoridentifier
Rights
- Language dc:language
- en
Identifiers
dc:identifier.*- Identifier
- oai:salford-repository.worktribe.com:5714768