Back to search

South Dakota State University

Study of an Automatic Sphygmomanometer System

Abstract

dc:description.abstract

A simple and dependable method of automatically measuring blood pressure indirectly would be of great help to the medical profession. An electronic system to implement the measurement of blood pressure using a double cuff method is presented. The scheme is to determine diastolic pressure from the amplitude of the difference signal obtained by subtracting the pulse wave observed in one cuff from that of the other, and to determine systolic pressure from the amplitude of the pulse wave observed in the cuff furthermost from the heart. Although the system is electronically practical, the results obtained from this system show that the difference signal does not give a dependable indication of the diastolic pressure point. The method of determining systolic pressure is workable.

Degree

thesis:*
Name thesis:degree_name
Master of Science (MS)
Level thesis:degree_level
Thesis - Open Access
Discipline thesis:degree_discipline
Electrical Engineering

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Morgan, James B.

Identifiers

dc:identifier.*
Repository record dc:identifier
https://openprairie.sdstate.edu/etd/3743

Chain of custody

source
Harvested from
South Dakota State University
Base URL
openprairie.sdstate.edu/do/oai/
Last updated
2026-11-19
Source record
OAI-PMH GetRecord
citation

Morgan, James B.. Study of an Automatic Sphygmomanometer System. Thesis - Open Access thesis, https://openprairie.sdstate.edu/etd/3743