Functional electrical stimulation based training orthosis for hand function following stroke

Abstract

dc:description.abstract

<p>Two types of device were developed. The first based on shoulder mounted switches, designed to be immune to misalignment by the user. The second type of device used the electromyogram (EMG) signal from the muscle that was being stimulated to control the stimulation. A novel method of removing the stimulation and M wave artefact was developed enabling the intensity of the stimulation to be controlled by the EMG envelope. Additionally, the stimulation could be triggered by EMG activity or maintained at a fixed level for as long as EMG activity was maintained. The devices were tested in a series of case studies and factors effecting their uses explored. In particular it was noticed that the effort in using a device directly impacted on the use of the device, increasing spastic tone and resisting the opening of the hand. In general, shoulder mounted switches had least effect on tone followed by EMG triggered stimulation, then EMG maintained stimulation and finally EMG proportional stimulation. It was observed that voluntary EMG activity was modulated by the stimulation and this effect was investigated further.</p> <p>From a review of the literature a model of the principle peripheral nervous system motor control structure was developed. The model was expressed in a series of linear equations and the effect of electrical stimulation examined. A series of experiments were devised to examine the effect of electrical stimulation on voluntary EMG. An apparatus for measuring isometric wrist torque while recording extensor and flexor EMG and a data acquisition system was constructed. Measurements with subjects with normal neurology confirmed the effect in the extensors and additionally in the flexor muscle groups. Graded voluntary effort and stimulation intensity tests enabled the main neurological processes to be identified and the majority of effects predicted by the model were observed.</p>

Degree

thesis:*
Name dc:type.qualificationname
Ph.D.
Level dc:type.qualificationlevel
doctoral
Grantor dc:publisher.institution
University of Southampton
Year dc:date.issued
2004

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Taylor, Paul Nicholas

Rights

Language dc:language
English, eng

Chain of custody

source
Harvested from
University of Southampton
Base URL
eprints.soton.ac.uk/cgi/oai2
Last updated
2026-11-19
Source record
OAI-PMH GetRecord
related terms
citation

Taylor, Paul Nicholas. Functional electrical stimulation based training orthosis for hand function following stroke. doctoral thesis, University of Southampton, 2004.