Back to search

University of Southampton

Numerical study of transient response in amperometric and potentiometric glucose biosensors

Abstract

dc:description.abstract

Biosensors are indispensable tools for continuous monitoring of key analytes, such as glucose, in chronic disease management. This thesis focuses on enzyme-immobilized redox polymer biosensors, leveraging reaction-diffusion partial differential equations (PDEs) to bridge the gap between theory and practical applications for biosensor design.<br/><br/>A transient numerical scheme in MATLAB was used to solve coupled non-linear PDEs, capturing the dynamic behavior of enzyme-membrane systems. Verification was achieved by comparing steady-state analytical solutions with numerical outputs, while Case diagrams were used to identify different kinetic and transport regimes. Experiments with deflavinated enzymes provided a means to validate our model by systematically manipulating key parameters. Simulation results were compared with experimental data, revealing key insights into transient enzyme-membrane interactions.<br/><br/>The study also investigated tubular flow electrodes, revealing the dependence of steady-state amperometric responses on solution viscosity under flow conditions. Experimentally obtained chronoamperograms were fitted to numerically solved dynamic simulations, confirming the model's predictive capabilities. Novel boundary conditions were also proposed to simulate potentiometric biosensors, broadening the scope of this work beyond amperometric sensing.<br/><br/>Animations visualizing concentration profiles and system responses offer a deeper understanding of transient and steady-state behaviors. These insights can contribute to the optimization of biosensor performance, including sensitivity, dynamic range, and response time, advancing the development of robust and accurate glucose monitoring technologies.

Degree

thesis:*
Name dc:type.qualificationname
Ph.D.
Level dc:type.qualificationlevel
doctoral
Grantor dc:publisher.institution
University of Southampton
Year dc:date.issued
2025

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Khan, Muhammad Hashim
Advisors dc:contributor.advisor
  • Bartlett, Philip
  • Denuault, Guy
ORCID person:identifier.orcid

Rights

Language dc:language
English, eng

Chain of custody

source
Harvested from
University of Southampton
Base URL
eprints.soton.ac.uk/cgi/oai2
Last updated
2026-11-19
Source record
OAI-PMH GetRecord
related terms
citation

Khan, Muhammad Hashim. Numerical study of transient response in amperometric and potentiometric glucose biosensors. doctoral thesis, University of Southampton, 2025.