Automatic Alignment Methods for VSM Using Magnetic Samples with Reduced Time and Higher Precision

Abstract

dc:description.abstract

Vibrating Sample Magnetometers (VSM) are used to measure the magnetic properties of ferro or ferrimagnetic materials. They have a wide range of users in both industry and in universities. Currently VSMs require a manual alignment for each sample loaded in the instrument. If the sample is not centered within the pickup coils, systematic errors are introduced into the data. Centering the sample requires significant time to complete and needs to be repeated after each field angle, field or temperature change. For this thesis project an automatic alignment system was designed, implemented and tested. The system measures the position of the sample after each field angle change and corrects for sample shifts after a field angle change avoiding large systematic errors in the My signal.

Degree

thesis:*
Name thesis:degree_name
Master of Science
Level thesis:degree_level
Masters
Discipline thesis:degree_discipline
Physics
Department thesis:degree_department
Physics
Grantor
Texas State University
Year dc:date.issued
2021

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Dieckow, Kolton J.
Advisor dc:contributor.advisor
  • Geerts, Wilhelmus Johannes Maria Arnoldus
Committee members dc:contributor.committeemember
  • Samwel, Erik
  • Smith, Casey
  • Miyahara, Yoichi

Subjects

dc:subject × 3

Rights

Language dc:language.iso
en

Identifiers

dc:identifier.*
Dc Identifier Citation
Dieckow, K. J. (2021). Automatic alignment methods for VSM using magnetic samples with reduced time and higher precision. [Master's thesis, Texas State University].

Chain of custody

source
Harvested from
Texas State University
Base URL
digital.library.txst.edu/server/oai/request
Last updated
2026-11-20
Source record
OAI-PMH GetRecord
citation

Dieckow, Kolton J.. Automatic Alignment Methods for VSM Using Magnetic Samples with Reduced Time and Higher Precision. Masters thesis, Texas State University, 2021. https://hdl.handle.net/10877/13528