Back to search

University of Texas at Austin

A mathematical model for simulation of solute transport within the patient-artificial kidney system

Abstract

dc:description.abstract

The advent of the Middle Molecule Theory of Uremic Toxicity has introduced the need for hemodialyzers capable of efficiently removing metabolites with molecular weights between 300 and 2000. The rate at which these solutes can be removed from the body depends not only upon the efficiency of the dialyzer, but also upon the ability of the metabolites to diffuse from innerbody compartments into the patient's blood stream. In an effort to characterize innerbody transport during hemodialysis, a multicompartmental patient-artificial kidney model has been developed. In vivo clinical data in the form of solute-plasma concentration decline measurements following an impulse IV injection of Dextran-1500 (1500 molecular weight dextran fraction) and Vitamin B-12 have been obtained. These data were used to determine model parameters in the form of compartmental volumes and intercompartmental mass transfer coefficients. Analysis of the dextran data indicated that two intracellular pools, the interstitial pool, and the plasma pool were involved in DX-1500 solute kinetics. Further analysis of the data yielded a transcapillary mass transfer coefficient of 502 ml/min and two transcellular coefficients of 141 and 7 ml/min. The Vitamin B-12 data produced a similar compartmental arrangement. A transcapillary mass transfer coefficient of 470 ml/min and two transcellular coefficients of 30 ml/min and 5 ml/min were obtained. The DX-1500 parameters were used to simulate various modes of dialysis. The model predicted that innerbody mass transfer barriers begin to limit solute removal even for moderate dialyzer clearances. In contrast to this, the model predicted that innerbody mass transfer has little effect on the dialysis of urea.

Degree

thesis:*
Name thesis:degree_name
Master of Science in Engineering
Level thesis:degree_level
Masters
Discipline thesis:degree_discipline
Mechanical Engineering
Department thesis:degree_department
Mechanical Engineering
Grantor
University of Texas at Austin
Year dc:date.issued
1974

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Hlavinka, Dennis J., 1946-
Advisor dc:contributor.advisor
  • Not available

Subjects

dc:subject × 4

Rights

dc:rights.accessrights
Statement dc:rights.accessrights
  • Open
  • Copyright © is held by the author. Presentation of this material on the Libraries' web site by University Libraries, The University of Texas at Austin was made possible under a limited license grant from the author who has retained all copyrights in the works.
Language dc:language.iso
eng

Identifiers

dc:identifier.*

Chain of custody

source
Harvested from
University of Texas
Base URL
repositories.lib.utexas.edu/server/oai/request
Last updated
2026-11-20
Source record
OAI-PMH GetRecord
citation

Hlavinka, Dennis J., 1946-. A mathematical model for simulation of solute transport within the patient-artificial kidney system. Masters thesis, University of Texas at Austin, 1974. https://doi.org/10.15781/T23N20X9R