University of Toronto
On the Development of a 2MHz Radial Imaging Ultrasound Array for Potential Use in Guiding Pedicle Screw Insertion
Abstract
dc:description.abstractSpinal fusion surgeries often require the insertion of screw implants into the pedicle bone. Although the procedure is somewhat routine, there is an inherent error rate when surgeons employ manual feedback techniques in the operating room. An alternative image guidance technique to reduce the error rate is an ultrasound transducer, which could be incorporated with the surgical toolkit to image the pedicle bone from within. This allows for real-time, non-radiating visualization of the anatomy and bone boundaries relative to the tip of the toolkit. The primary objective of this thesis is to describe the design and fabrication of 2MHz ultrasound probes for pedicle screw guidance. Three single-element transducer designs (3.5mm diameter) are explored in order to arrive at an acoustic design (backing, active, matching layer) that enables intra-pedicular imaging. To eliminate the need for manual rotation of the transducer, we have successfully designed and fabricated a 2MHz radial array that consists of 32 elements (3.2mm outer diameter). Using a commercial platform, we have obtained radial images from successive groupings of four array elements, when the array was placed in various fluids. Although this work has not demonstrated intra-pedicular imaging, we believe that it represents an important step in this direction. A secondary objective of this work is to investigate the directionality distribution in the pedicle's trabecular bone. This stems from the need to better understand the manner in which it can affect ultrasound propagation. In this thesis, we have estimated the direction distribution of pedicle trabecular bone from micro-CT images using Gabor filters. The Gabor filter was tested for suitability in detecting trabecular structures first through simulated 2-D line tests then through 3-D bone models. When applied to micro-CT images of six human pedicle bones, the final results are an estimation of the main directions of trabecular structures in both the coronal and sagittal planes. However, since the pedicles tested were from just one cadaver, more general conclusions cannot be drawn concerning the directionality distribution. Nonetheless, a reliable quantitative method has been developed that should have considerable potential in a variety of applications.
Degree
thesis:*- Department dc:contributor.department
- Biomedical Engineering
- Year dc:date.issued
- 2015
Author and committee
dc:creator, dc:contributor.*- Author dc:creator
-
- Manbachi, Amir
- Advisors dc:contributor.advisor
-
- Cobbold, Richard S.C.
- Ginsberg, Howard J.
Subjects
dc:subject × 6Identifiers
dc:identifier.*- Handle dc:identifier.uri
- https://hdl.handle.net/1807/69430