Abstract
dc:description.abstractFor imaging nano-scaled samples, atomic force microscopy (AFM) and scanning electron microscopy (SEM) are two complementary techniques that have become indispensable in nano research. A hybrid SEM-AFM instrument can enable in-situ complementary imaging, fast sample navigation down to nanometer scale, SEM-guided nanoindentation, live monitoring of AFM tip condition, and visually guided nanomanipulation. A closed-loop operated AFM system was developed for use within an SEM, which would significantly benefit nano-structure and nano-materials research. Electron irradiation from SEM induces significant fluctuations in the AFM force sensor readout, which compromises the AFM functionality. To enable truly simultaneous SEM-AFM imaging, an active drift reduction method was proposed to overcome this problem, which is based on the junction field-effect transistor (JFET) model of the piezoresistive AFM cantilever. The experiment results showed that the proposed method was effective, reducing the drift by about 2000 times down to
Degree
thesis:*- Department dc:contributor.department
- Mechanical and Industrial Engineering
- Year dc:date.issued
- 2018
Author and committee
dc:creator, dc:contributor.*- Author dc:creator
-
- Chen, Jun
- Advisor dc:contributor.advisor
-
- Sun, Yu
Subjects
dc:subject × 6Identifiers
dc:identifier.*- Handle dc:identifier.uri
- https://hdl.handle.net/1807/99973