Finite Element Analysis Of Post Vs Threaded-Post Glenoid Baseplates In Reverse Total Shoulder Arthroplasty: A Comparative Study

Abstract

dc:description.abstract

This study compared a standard‑post baseplate with a threaded‑post baseplate in reverse total shoulder arthroplasty (rTSA) to determine whether threaded central fixation improves micromotion and stress distribution. A finite element model of a left scapula with a commercially available rTSA system was created; the two designs differed only by the central post (standard press‑fit versus threaded). A 756 N physiological load was applied at 60°, 90°, and 120° of abduction. Outcomes included total and directional relative micromotion at the bone–implant interface and von Mises stress in the glenoid and baseplate. Total micromotion remained below the 150 µm osseointegration threshold for both designs. At 60° and 120°, the threaded-post significantly reduced mean total micromotion (60°: 13.80 µm versus 18.37 µm, p < 0.001; 120°: 15.41 µm versus 16.46 µm, p < 0.001). Directional displacements were also lower with the threaded-post, especially lateral pull‑away (60°: 1.62 µm versus 7.2 µm; 120°: 0.53 µm versus 9.5 µm). Stress in the glenoid and baseplate was consistently lower and more uniformly distributed with the threaded design. At 90°, no significant differences were observed. The threaded‑post baseplate provided superior fixation under shear‑dominant loading, reducing micromotion and stress while maintaining more uniform load distribution. These findings support the use of a threaded central post to enhance initial stability and potentially improve long‑term osseointegration in rTSA; however, the threaded post may be best reserved for patients with high functional demands, poor bone quality, or when revision arthroplasty is anticipated, as the added operative time and technical precision required for rotational alignment may outweigh its biomechanical benefits in lower‑risk patients.

Degree

thesis:*
Name thesis:degree_name
Master of Science in Biomedical Engineering (MSBME)
Level thesis:degree_level
Thesis
Discipline thesis:degree_discipline
Biomedical Engineering

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Morgan, Christopher
Contributors dc:contributor
  • <p>Alan Eberhardt</p>
  • <p>David Littlefield<br>Eugene Brabston</p>

Identifiers

dc:identifier.*

Chain of custody

source
Harvested from
University of Alabama Birmingham
Base URL
digitalcommons.library.uab.edu/do/oai/
Last updated
2026-11-19
Source record
OAI-PMH GetRecord
related terms
citation

Morgan, Christopher. Finite Element Analysis Of Post Vs Threaded-Post Glenoid Baseplates In Reverse Total Shoulder Arthroplasty: A Comparative Study. Thesis thesis, https://digitalcommons.library.uab.edu/etd-2020s/231