Experimental Determination of Shielding Requirements for PET Medical Facilities

Abstract

dc:description.abstract

Purpose: Shielding considerations are important for designing a positron emission tomography/computed tomography (PET/CT) imaging facility, because of the high energy (511 keV) of electron-positron annihilation photons. Since radiation emitted from patients administered PET radiopharmaceuticals includes lower-energy scattered photons, a given thickness of lead attenuates more of the radiation than it would for a monoenergetic beam of 511 keV photons. A more accurate determination of the effective attenuation coefficient of lead for the spectra of photon energies emitted from PET/CT patients could reduce the shielding requirements necessary to adequately protect public health. In this work, the spectra of energies emitted from 14 patients of different Body Mass Index (BMI) are presented. Methods: The radiation spectra emitted from adult patients injected with F-18 FDG at The Kirklin Clinic PET/CT Center were measured through an unleaded wall, and a wall that contains 15.9 mm of lead. The spectra were corrected for radioactive decay time, distance from the patient, detection efficiency of the sodium iodide detector at different energies, and normalized to an injected activity of 15 mCi. The counts were summed to produce a measure of the quantity of transmitted radiation. The shielded wall sums were divided by the unshielded wall sums to determine the radiation transmission factor for each patient. Results: An effective linear attenuation coefficient of c = 0.226/mm was measured in this work. The results were not significantly affected by differences in patient BMI. For a particular lead shielding thickness, new broad-beam transmission factors calculated using c = 0.226/mm are less than the corresponding broad-beam transmission factors in lead at 511 keV given by American Association of Physicists in Medicine Task Group 108. Conclusions: Since much of the 511 keV annihilation radiation is scattered in the human body, a spectrum of photon energies interacts with the lead shielding. This results in an increase in the effective attenuation coefficient for lead compared to a monoenergetic beam of 511 keV photons. Because of these effects, current shielding designs for PET/CT imaging facilities using lead shielding may result in safety margins that are higher than originally calculated to adequately protect public health.

Degree

thesis:*
Level thesis:degree_level
Dissertation

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Brinkley, Bradley Scott
Contributors dc:contributor
  • Claudiu T Lungu
  • Alfred A Bartolucci
  • Steven M Becker
  • Riedar K Oestenstad
  • Sharon L White
  • Michael V Yester

Subjects

dc:subject × 7

Identifiers

dc:identifier.*

Chain of custody

source
Harvested from
University of Alabama Birmingham
Base URL
digitalcommons.library.uab.edu/do/oai/
Last updated
2026-11-19
Source record
OAI-PMH GetRecord
citation

Brinkley, Bradley Scott. Experimental Determination of Shielding Requirements for PET Medical Facilities. Dissertation thesis, https://digitalcommons.library.uab.edu/etd-collection/1259