Validation of oscillometric blood pressure measuring devices : a case study of the BpTRUtm

Abstract

dc:description

Hypertension is one of the most common reasons why North Americans visit a physician's office and its measurement technique has been reasonably standard since the earliest accepted description of an indirect method of measuring blood pressure by Riva Rocci in 1896. Korotkoff later modified the traditional sphygmomanometer in 1905. However, since then the mercury sphygmomanometer has become not only the gold standard, but also an essential diagnostic tool in everyday medical care. Recently however, there has been growing concern over the accuracy of the measurements obtained and the potential biohazard with standard mercury sphygmomanometers. Many cities and countries in Europe and North America, are in the process of phasing them out. Newer instruments such as the auscultatory aneroid, and the automated electronic devices of which the majority are of the oscillometric type, are beginning to appear. All new devices have the option of undergoing a validation process to assess the accuracy according to the relevant governing bodies. The BpTRU™ is an automated oscillometric electronic blood pressure device developed by VSM Medtech Ltd of Vancouver. Connected in parallel by means of a T-tube, to the current gold standard mercury sphygmomanometer, allowed blood pressures to be measured simultaneously. Two observers blinded from each other and the device, recorded blood pressures individually. The mean observer average was compared to the device mean blood pressures for agreement. This was done according to the American National Standard for Electronic or Automated Sphygmomanometers-1992 and the British Hypertension Society protocol-1993. The validation process was initially conducted in adults and agreement was within the above standards. However the device tended to underestimate at higher systolic blood pressures, so the algorithm was modified and the device re-validated using embedded raw data. After running the validation study in children, which again satisfied the above criteria, the raw data was combined in a final study. The BpTRU™ is the only automated blood pressure measuring device that has been independently validated and has attained a high level of accuracy over this broad age and blood pressure range.

Degree

thesis:*
Name thesis:degree_name
Master of Science - MSc
Level thesis:degree_level
master's
Discipline thesis:degree_discipline
Pharmacology and Therapeutics
Grantor dc:publisher
University of British Columbia
Year dc:date
2003

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Mattu, Gurdial Singh

Rights

dc:rights
Statement dc:rights
  • For non-commercial purposes only, such as research, private study and education. Additional conditions apply, see Terms of Use https://open.library.ubc.ca/terms_of_use.
Language dc:language
eng

Identifiers

dc:identifier.*
Handle dc:identifier
http://hdl.handle.net/2429/14252

Chain of custody

source
Harvested from
University of British Columbia
Base URL
circle.library.ubc.ca/oai/request
Last updated
2026-11-19
Source record
OAI-PMH GetRecord
related terms
citation

Mattu, Gurdial Singh. Validation of oscillometric blood pressure measuring devices : a case study of the BpTRUtm. master's thesis, University of British Columbia, 2003. http://hdl.handle.net/2429/14252