Back to search

University of British Columbia

Intensity modulation of therapeutic photon beams using a rotating multileaf collimator

Abstract

dc:description

This thesis describes the development and implementation of a novel method of delivering intensity modulated radiation therapy (IMRT) that provides greater accuracy and spatial resolution than currently available methods. Through improvements in multileaf collimator (MLC) based fluence generation, a dose distribution may be generated that conforms more closely to the tumour target volume. Healthy tissue surrounding the target volume will therefore receive less dose, reducing the probability of side effects and allowing the physician to increase the prescribed tumour dose (dose escalation). MLC based IMRT techniques are well established but suffer several physical limitations. Dosimetric spatial resolution is limited by the MLC leaf width, interleaf leakage and tongue-and-groove effects degrade dosimetric accuracy and the range of leaf motion limits the maximum deliverable field size. Based on observations from a linear systems model of dosimetric spatial resolution degradation it is hypothesized that, by rotating the entire MLC between each sub-field, improvements will be obtained in spatial resolution, dosimetric accuracy and maximum deliverable field size. To generate arbitrary fluence maps in this way, a series of unique algorithms were developed that are capable of determining the necessary rotated MLC segments. These IMRT fields may be delivered statically (with the collimator rotating to a new position in between sub-fields) or dynamically (with the collimator rotating and leaves moving simultaneously during irradiation). A full description of the rotational leaf motion algorithms is provided. An analysis of the rotating leaf motion calculation algorithms with focus on radiation efficiency, the range of collimator rotation and number of segments is provided. The mechanical and radiation producing characteristics of standard linear accelerators under collimator rotation conditions are also investigated. The technique is evaluated by characterizing the ability of the algorithms to generate rotating leaf sequences for desired fluence maps. Comparisons are also made between our method and conventional sliding window and step-and-shoot techniques. Results show improvements in spatial resolution, reduced interleaf effects and maximum deliverable field size over conventional techniques. Clinical application of these enhancements can be realized immediately with static rotational delivery although improved control of the MLC will be required for dynamic delivery.

Degree

thesis:*
Name thesis:degree_name
Doctor of Philosophy - PhD
Level thesis:degree_level
doctoral
Discipline thesis:degree_discipline
Physics
Grantor dc:publisher
University of British Columbia
Year dc:date
2003

Author and committee

dc:creator, dc:contributor.*
Author dc:creator
  • Otto, Karl

Rights

dc:rights
Statement dc:rights
  • For non-commercial purposes only, such as research, private study and education. Additional conditions apply, see Terms of Use https://open.library.ubc.ca/terms_of_use.
Language dc:language
eng

Identifiers

dc:identifier.*
Handle dc:identifier
http://hdl.handle.net/2429/14891

Chain of custody

source
Harvested from
University of British Columbia
Base URL
circle.library.ubc.ca/oai/request
Last updated
2026-11-19
Source record
OAI-PMH GetRecord
related terms
citation

Otto, Karl. Intensity modulation of therapeutic photon beams using a rotating multileaf collimator. doctoral thesis, University of British Columbia, 2003. http://hdl.handle.net/2429/14891